![Inogen One G5 User Manual Download Page 24](http://html1.mh-extra.com/html/inogen/one-g5/one-g5_user-manual_2064306024.webp)
21
10. Push the column assembly into the
device such that the columns are
fully seated into the Inogen One® G5
concentrator. The spring loaded latch
button should fully return to the closed
position.
11. Connect the AC power supply cord to
the Inogen One® G5 and plug the power
supply AC cord into an electrical outlet.
Do Not Power on the Inogen One® G5
concentrator.
The following steps can be done by
pressing specific buttons on the device’s
screen or within Inogen Connect App.
Steps from your device’s display:
º
Press and hold the plus (+) and (-)
minus button for 5 seconds. The
screen will display the following
informational icon. Release button
once icon is displayed on screen.
º
Press the alert button once and
screen will display the following
informational icons.
º
Press the power button to turn
on the Inogen One® G5, and use
normally.
Steps using Inogen Connect App.:
º
If you are using Inogen Connect
App, navigate to the Advanced
screen, then to Additional
Information screen and click on
the Column Reset button.
Closed and locked
10