![Inogen One G5 User Manual Download Page 18](http://html1.mh-extra.com/html/inogen/one-g5/one-g5_user-manual_2064306018.webp)
15
HH:MM
Vx.x:SN
O2
!
C
Notification Icons (Continued)
Description
Display Icons
Condition/Action/Explanation
Please Wait
Shutting Down
On/Off button has been pressed for two
seconds. Concentrator is performing
system shut down.
Hours:Minutes
Software Version:
Serial Number
Audible Alert button has been pressed for five
seconds.
Low Priority Alerts
The following low priority alerts are accompanied by a double beep and a solid yellow light.
Description
Display Icons
Condition/Action/Explanation
Battery Low
Attach Plug
Battery power is low, with less than 10
minutes remaining. Attach external power
supply or power down and insert a fully
charged battery.
Replace Columns
Column maintenance is required within 30 days.
Contact your equipment provider to arrange for
service.
Check Battery
Battery error has occurred. Check the
connection of your battery and ensure that it is
properly attached and latched on concentrator.
If battery error recurs with same battery, stop
using the battery and switch to a new battery or
remove battery and operate concentrator using
external power supply.
Oxygen Low
Concentrator is producing oxygen at a
slightly low level (<82%) for a period of
10 minutes. If condition persists, contact
your equipment provider.
Remove Battery
to Cool
Battery has exceeded its charging temperature
and charging has stopped. The battery will not
charge while this alert is present but will begin
to charge when the battery temperature
returns to the normal operating range. If
battery charging is desired sooner, remove the
battery from the concentrator and allow it to
cool in an open area for approximately 10-15
minutes. Then, re-insert the battery into the
Inogen One® G5. If the problem still persists,
contact your equipment provider.
Summary of Contents for One G5
Page 1: ...1 USER MANUAL ...
Page 2: ...2 ...