Condition/Action/Explanation
Battery has exceeded temperature limit while concentrator
is running on battery power. If possible, move concentrator
to a cooler location or power unit with an external power
supply and remove battery. If condition persists, contact
your equipment provider.
Concentrator is not detecting a breath. Check that:
cannula is connected to concentrator, there are no kinks
in tubing, and cannula is positioned properly in your nose.
If Satellite Conserver is attached, make sure flow control
on concentrator is set to Satellite Mode. This alert may be
enabled or disabled through use of the mode button.
Concentrator is experiencing an error, but is capable of
continued operation. As soon as you are able:
• Remove and re-insert battery, and/or
• Check external power supply connections.
If condition persists, contact your equipment provider.
Oxygen output concentration is below 50%. If condition
persists, switch to your backup oxygen source and contact
your equipment provider to arrange for service.
Message Display & Text
Battery HOT
Warning
No Breath Detect
Check Cannula
System Error
See Manual
Oxygen Error
Service Needed
Medium Priority Alerts
The following medium priority alert messages are accompanied by a
triple beep
,
repeated every 25 seconds, and a
flashing red light
.
26
Condition/Action/Explanation
Concentrator has insufficient battery power to produce
oxygen. Attach external power supply or exchange battery,
then restart unit if necessary by pressing On/Off button.
Battery has exceeded temperature limit while concentrator
is running on battery power. Concentrator has stopped
producing oxygen. If possible, move concentrator to a
cooler location, then turn power off and back on. Ensure air
intake and outlet vents have clear access and particle filter
is clean. If condition persists, switch to a backup source of
oxygen and contact your equipment provider.
Concentrator temperature is too high, and oxygen
production is shutting down. Ensure air intake and outlet
vents have clear access and particle filter is clean. If
condition persists, switch to a backup source of oxygen and
contact your equipment provider.
Message Display & Text
Battery Empty
Attach Plug
Battery HOT
Shut Down
System HOT
Shut Down
High Priority Alerts
CAUTION
If you are not near the Inogen One® G2 you may not be able to hear or
see the high priority alerts. Make sure the Inogen One® G2 is in a
location where the alerts and alarms will be recognized if they occur.
The following high priority alert messages are accompanied by a
five beep pattern
,
repeated every 10 seconds, and a
flashing red light
.
Audible and Visible Signals Chapter 4
27