5. ADVANCED TESTS
RAMCHECK LX OWNER’S MANUAL
speed limit.
There are two modes for presetting speed. The default mode is "SPEED AT"
where the RAMCHECK LX speed is actually fixed at the preset speed. This
mode is useful when the actual speed is unimportant, so long as it meets a
particular reference value. The other mode is the "SPEED LIMIT" where
RAMCHECK LX fails a unit that is slower than the preset speed limit. If the
unit tests slower than the preset speed, the test terminates with a "SPEED
ERROR: ABOVE LIMIT" message. If the unit tests faster than the preset
limit, the actual speed is displayed. Note that if you select the SPEED LIMIT
mode, the preset speed will be shown with a “ ^ ” prefix during all subsequent
test modes.
SPEED
ALGORITHM setup
is only used by legacy
EDO/FPM devices.
5.3.2.2 SPEED ALGORITHM SETUP
If you select AUTO SPEED, or if you select a certain speed value to be used
as a limit, you will be prompted to select the speed algorithm.
This will determine the resolution at which RAMCHECK LX will make its
speed determination. The broader the resolution, the quicker the test. For a
more precise speed measurement, select the default by pressing
F1
.
PATTERNS SETUP:
→
→
→
5.3.4 SETUP PATTERNS
This selection (from the second menu of SETUP PARAMETERS) allows you
to edit two of the main 32-bit test patterns used by BASIC test and similar
tests. Default patterns are PATT1= 55555555h, PATT2=AAAAAAAAh.
Setup uses hexadecimal notation.
TESTFLOW
SETUP:
→
5.4 TESTFLOW
Press
F2
from the main setup menu to enter the testflow setup. This menu
will allow you to setup your own customized program test flow.
Pressing
F4
from this menu resets the program to its default testflow. The
following display appears upon resetting the test flow to its default mode:
PAGE 28