![Innovative Neurological Devices Cervella User Manual Download Page 33](http://html1.mh-extra.com/html/innovative-neurological-devices/cervella/cervella_user-manual_2063938033.webp)
PAGE 33 OF 38
Guidance and manufacturer´s declaration – electromagnetic emission
Cervella
is intended for use in the electromagnetic environment specified below. The customer or
the user of
Cervella
should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
Cervella
uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B
Cervella
is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-
voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Not Complies
Voltage fluctuations
flicker emissions
IEC 61000-3-3
Complies