www.innovatest-europe.com                                 7

Table 1:

  Measuring ranges   Measuring ranges 

 

(Ra) 

  (Rz) 

  Cut-off 

(lc)

  0.03µm Ra 0.1µm 

0.1µm Rz 0.5µm   

0.25mm

 0.1µm<Ra<2.0µm  0.5µm<Rz<10.0µm 

0.8mm

  2.0µm Ra 6.3µm 

10.0µm Rz 50.0µm 

2.5mm

The tester can be switch on with the “>” (start) key and be turned off automatically after 

2 minutes.

If the tester is auto switched off, the last measuring result can’t be shown on the screen 

when re-start the tester as this auto-off is the real off (power cut-off).

Summary of Contents for R130

Page 1: ...MANUAL TH 170 Universal hardness tester MANUAL TH 170 Universal hardness tester MANUAL R130 R135 Roughness tester ...

Page 2: ...e due to new technologies and continuous development We reserve the right to change or modify specifications of products without prior notice We recommend you to contact our sales office for up to date information All rights reserved 2 www innovatest europe com ...

Page 3: ...and digital display 6 6 Calibration 8 7 Operation 10 8 Probes 11 8 1 General purpose probe 11 8 2 Transverse chisel probe 11 8 3 Parallel chisel probe 11 8 4 Small bore probe 12 8 5 Groove bottom probe 12 9 Probe replacement 13 10 Analog signals output 14 11 Maintenance 15 11 1 Protection storage 15 11 2 Cleaning 15 11 3 Repairs 15 12 Accessory height stand 16 ...

Page 4: ...surface roughnesss testing is required NOTE Even though this instrument is designed and built to withstand the rigors of handling and use it is a precision instrument and should be treated with care to assure measurement accuracy and reliable performance R130 standard kit is furnished in a fitted case and includes R130 unit with a general purpose probe R130 2002 reference specimen 9 volt alkaline ...

Page 5: ...averse length Evaluation length Numbers of cutoff 0 75mm 0 25mm 1 1 25mm 0 75mm 3 1 75mm 1 25mm 5 Cut off 0 8mm Traverse length Evaluation length Numbers of cutoff 1 2mm 0 8mm 1 3 0mm 2 4mm 3 4 5mm 4 0mm 5 Cut off 2 5mm Traverse length Evaluation length Numbers of cutoff 3 0mm 2 5mm 1 Probe type Piezoelectric Maximum stylus force 15 0mN 1500mgf Power 9 volt consume type alkaline battery Battery ca...

Page 6: ...ameters Ra Roughness Average the arithmetic average height of roughness irregularities measured from a mean line within the evaluation length L Rz Maximum Roughness Depth The largest of the peak_to_valley roughness depths in the evaluation length ...

Page 7: ... 2 Start button 3 Parameter selection button 4 Cut off selection switch 5 Traverse length switch 6 Meter inch switch 7 Output connector accessory 9 Battery alkaline 10 Battery compartment cover 11 Protective cover 12 2 V feet 13 Probe holding block 14 Probe ...

Page 8: ...Set Cut off selection switch for 0 25mm 0 8mm 2 5mm base on the roughness value be estimated Refer to Table 1 Set traverse length switch for 1 3 or 5 Refer to specification section s traverse length for detail 3 Press and release start button the readout should display 1 X XX Ra If the readout display 8 888 it indicating a low battery and need to replace the battery 4 If the Ra or Rz value display...

Page 9: ...1µm Ra 2 0µm 0 5µm Rz 10 0µm 0 8mm 2 0µm Ra 6 3µm 10 0µm Rz 50 0µm 2 5mm The tester can be switch on with the start key and be turned off automatically after 2 minutes If the tester is auto switched off the last measuring result can t be shown on the screen when re start the tester as this auto off is the real off power cut off ...

Page 10: ...en If the reading is within 0 1µm of the value stated on the report of calibration certificate calibration is within tolerance If the reading differs from the value stated on the report of calibration certificate by more than 0 1µm take additional readings around the central area of the specimen If the reading still differ by more than the allowable tolerance recalibrate the tester following the p...

Page 11: ...n calibrating with a Small Bore Probe or Groove Bottom Probe the probe should be in the 180 position And make sure that the tester base is on the same height as the surface of the calibration specimen calibration plate or height stand ...

Page 12: ... not to be used for measurement or tester performance evaluation purposes The tester can be hand held or placed on a surface in any attitude it will operate in virtually any position horizontally vertically at any angle in between even upside down With the probe positioned on the workpiece surface to be measured carefully adjust the tester and workpiece setup so that the Reference Line the upside ...

Page 13: ...ical diamond stylus 0002 5µm radius per DIN standards 8 2 Transverse chisel probe R130 2003 For gaging sharp edges or small O D s where probe is aligned with in 180 or closed position to axis of traverse 90 diamond chisel stylus 0004 10µm radius 8 3 Parallel chisel probe R130 2004 For gaging sharp edges or small O D s where probe is perpendicular in 90 or 270 position to axis of traverse 90 diamon...

Page 14: ...r of 5 0mm up to a depth of 15 0mm 90 conical diamond stylus 0004 10µm radius for R130 2005 0002 5µm radius for R130 2006 8 5 Groove bottom probe R130 2007 For measuring the bottoms of O ring grooves recesses and holes to depth of 6 0mm Also used for short lands and shoulders 90 conical diamond stylus 0004 10µm radius ...

Page 15: ...ockand the knurled locking collar Do not grasp the probe by its skid and stylus end Do not remove the knurled locking collar from the probe mounting block loosen it only enough to slide the probe out easily 3 Reverse the procedure to install a probe aligning the pin on the bottom of the probe body with the acess solt in the locking collar Make sure that the probe is fully inserted into the probe m...

Page 16: ...tput feature permits measurement results to be transmitted to an external device such as a printer computer or data collection system For further information concerning the signals output please contact the supplier Printer settings TA 220 settings 1 3 5 on TA 230 settings 2 3 6 7 on ...

Page 17: ...se 11 2 Cleaning To clean the tester unit use a soft lint free cloth moistened with a mild non abrasive liquid or foam cleaning agent Using a magnifier periodically inspect the probe skid and stylus area for dust dirt or other contamination To clean the probe skid and stylus use a soft camel s hair artist s brush moistened with denatured alcohol 11 3 Repairs When finded any trouble in using the te...

Page 18: ...s with the tapped hole in the bottom of the tester housing Tighten the screw by finger 2 Position the height Stand so that the probe is above the workpiece surface to be measured Use the coarse height adjustment to bring the probe skid and stylus barely in contact with the surface Use the fine adjustment to make the reference line the upside of the mounting block parallel to the bottom of the test...

Page 19: ... the manufacturing drawings which form part of this declaration Assessment of compliance of the product with the requirements relating to the compatibility was based on the following standards EN55022 EN60555 2 EN60555 3 EN50082 1 This declaration is the responsibility of the manufacturer importer Company name INNOVATEST Europe BV Address Borgharenweg 140 Telephone Telefax 31 43 3520060 31 43 3631...

Page 20: ... HEAD OFFICE INNOVATEST Europe BV MANUFACTURING DISTRIBUTION SERVICE Borgharenweg 140 6222 AA Maastricht The Netherlands Phone 31 43 3520060 Fax 31 43 3631168 E mail info innovatest europe com Website www innovatest europe com ...

Reviews: