www.innovatest-europe.com 3
2. Specifi cation
- Measuring ranges :
Ra--0.03µm~6.3µm/1µ”~250µ”
Rz--0.2µm ~18.5µm/8µ”~999µ”
- Display Resolution:
0.01µm /1µ”
- Cut-off:
0.25mm/0.8mm/2.5mm, ANSI 2RC Filter
- Display:
3-digital LCD
- Measure Accuracy:
Meets ISO and DIN standards
Traverse lengths and evaluation lengths:
Cut-off = 0.25mm
Traverse length
Evaluation length
Numbers of cutoff
0.75mm
0.25mm
1
1.25mm
0.75mm
3
1.75mm
1.25mm
5
Cut-off = 0.8mm
Traverse length
Evaluation length
Numbers of cutoff
1.2mm 0.8mm
1
3.0mm 2.4mm
3
4.5mm 4.0mm
5
Cut-off = 2.5mm
Traverse length
Evaluation length
Numbers of cutoff
3.0mm 2.5mm
1
- Probe type :
Piezoelectric
- Maximum stylus force: 15.0mN / 1500mgf
- Power:
9-volt consume-type alkaline battery
- Battery capacity:
Approx. 3000 measurements
- Operating temperature:10°~45°C / 50°~113°F
- Storage temperature : 0°~60°C / 32°~147°F