23
Manufacturer’s guide and declaration – electromagnetic immunities
INN-011 is intended to be used in the electromagnetic environment specified below. The patient or use of INN-
011 should assure it’s used in this environment. Portable and mobile RF communications equipment should not
be used closer to any part of, including cables, than the recommended separation distance calculated with the
equation applicable to the transmitter frequency.
Immunity test
IEC 60601 test level
Compliance
level
Recommended
separation
distance d:
Conducted RF
IEC 61000-4-6
3 Veff from 150kHz to 80 MHz
6 Veff in ISM amateurs bands between 150 KHz
and 80 MHz
3 Veff&6Veff
d= 30 cm
Irradiated RF
IEC 61000-4-3
10 V/m from 80 MHz to 2,7 GHz
10 V/m
d= 30 cm
Proximity field
immunity from wi-
reless RF commu-
nication devicesIEC
61000-4-3
TETRA 400
380 – 390 MHz
27 V/m
27 V/m
d= 30 cm
GMRS 460 FRS 460
430 – 170 MHz
28 V/m
28 V/m
LTE Band 13, 17
704 – 787 MHz
9 V/m
9 V/m
GSM 800/900, TETRA 800, iDEN 820,
CDMA 850, LTE Band 5
800 960 MHz
28 V/m
28 V/m
GSM 1800; CDMA 1900; GSM 1900;
DECT; LTE Band 5
1700 – 1990 MHz
28 V/m
28 V/m
Bluetooth, WLAN, 802.11 b/g/n, RIFD
2450, LTE Band 70
2400 – 2570 MHz
28 V/m
28 V/m
WLAN 802.11 a/n
5100 – 5800 MHz
9 V/m
9 V/m