Guidance and declaration of manufacturer – electromagnetic immunity
IINN-008 is intended to be used in the electromagnetic environment specified below. The customer of
user of INN-008 must assure it’s used in this environment
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment
Electrostatic
discharge (ESD) IEC
61000-4-2
+- 8 kV contact
+- 8kV contact
Floors must be made
of wood, concrete or
ceramic tiles. If the
floors are covered with
synthetic material, the
relative humidity should
be at least 30%
+- 2; 4; 8; 15 kV air +- 2; 4; 8; 15 kV air
Campo magnetico
ad alta frequenza
(50/60Hz) IEC
61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
GB