INHECO
Doc ID: 900905-000
October 2018
07
2
PrODUCT DESCrIPTION
2.1.
Intended Use
The OVT is designed for use as a verification tool to verify the temperature specifications
of fully functional ODTC
®
. Mechanical / Hardware and Firmware defects / problems
cannot be detected by the OVT. The OVT is not intended to be used to verify other
Thermal Cycler systems.
The OVT devices are delivered with CE- and UL- certification.
The OVT is designed specifically for use in Life Science and In Vitro Diagnostics. The
provided OVT software operates the OVT/ODTC
®
and evaluates the measurement data
of the OVT system and generates an official verification certificate
When using the OVT in a Biosafety Laboratory Environment, the user is responsible for
labeling the devices according to the WHO Laboratory Biosafety Manual
(ISBN 92 4154650 6) and for operating the devices according to this Biosafety Manual.
The OVT must be used exclusively by service technicians and user of workstation
system who are familiar with the instructions of this manual and the ODTC
®
Manual.
2.2.
Components - scope of supply
Before initial operation, make sure that the shipment of your unit and its scope of supply
is complete and no parts are damaged.
In case of parcel or product damages, take photos of the damaged boxes and products
and email them to
without delay. Transportation damages
must be reported to INHECO within 7 days of delivery. The following components should
be included in each shipment:
Fig.1: Components (measurement scale not correct)
( 1 ) OVT measuring Head and OVT Data Logger tightly connected with a cable (cable longer than shown in
the image)
( 2 ) USB flash drive with OVT Control Software and Data Logger driver
( 3 a ) USB cable ( 1.5m length )
( 3 b ) USB cable (3.0m length)
( 4 ) RJ45 Ethernet cable
( 5 ) Socket wrench ( 1.5 mm )
( 6 ) Certificate
( 7 ) Transport case
2
6
1
3a
3b
5
7
4