8-2
8-3
Figure 8.1. Location of Calibration Adjustments
8.4.1. AMPLIFIER OFFSET ADJUSTMENT
Select Mode I, select display of current mean value. Leave the current inputs open. Select 5A
input and 15mA current range.
Adjust Idc to obtain a minimal reading for the current mean value. The reading is typically
±
20uA.
8.4.2. CURRENT
CALIBRATION
Apply a 60Hz current to the 5A input „IN5A“ or the 30A input „IN30A“.Make sure you have
selected the correct current input on the Spectral Multimeter front panel.
Calibrator Output
Input
Range
Adjustment, rms current
2A (3A)
50mA
IN30A
3A
adjust I1 for 2.0000A (3.0000A)
IN5A 50mA adjust I2 for 50.000mA
Repeat above steps at least once before proceeding to the steps below.
15mA
150mA
1.5A
IN5A
IN5A 15mA adjust I3 for 15.000mA
150mA adjust for 150.00mA
IN5A 1.5A
adjust I5 for 1.5000A
Spectral M ultim eter Rear Panel
Am plifier
I6
I2
I5
I3
I4
I1
Idc
I7
8.4.3. VOLTAGE
CALIBRATION
Select Mode U. Select 20V range and apply 20V/60Hz. Adjust I6 for 20.000 Vrms display.
8.4.4. SHUNT INPUT CALIBRATION
Select Mode I. Select 600mV shunt range. Apply 600mV/60Hz and adjust I7 for 10.000 Arms
display.
8.4.5. POWER / FLICKER / BURST / CALIBRATION
Power is the precise product of current and voltage and is also calibrated with above steps. Flicker
and Burst is also calibrated with above steps.