POWer:ACTive
<
F
>
Query or set field DC coupled power in Watt
:AC
<
F
>
Query or set field AC coupled power in Watt
:INT
<
F
>
Query or set field DC coupled average energy (short time
integration)
:AC
<
F
>
Query or set field AC coupled average energy (short time
integration)
:APParent
<
F
>
Query or set field DC coupled apparent power
:AC
<
F
>
Query or set field AC coupled apparent power
6-9
6-10
:INT
<
F
>
Query or set field DC coupled average apparent energy
(short time integration)
:AC
<
F
>
Query or set field AC coupled average apparent energy
(short time integration)
:REActive
<
F
>
Query or set field DC coupled reactive power
:AC
<
F
>
Query or set field AC coupled reactive power
:INT
<
F
>
Query or set field DC coupled average reactive energy
(short time integration)
:AC
<
F
>
Query or set field AC coupled average reactive energy
(short time integration)
:FFT
<
F
>
:G
Set the field (0,1,...,9) for the power harmonic previously
selected by the FORMat:START and FORMat:END
command. Use the G argument instead of
<
F
>
to display
FFT(p) in the display graphic zone.
:FFT?
Query all power harmonics in the range specified by the
FORMat:STart and FORMat:END commands.
:
FACtor
<
F
>
Query or set the field for the DC coupled power factor
:AC
<
F
>
Query or set the field for the AC coupled power factor
ENergy:ACTive
<
F
>
Query or set field of energy (long time integration)
:APParent
<
F
>
Query or set field of apparent energy (long time
integration)
:REActive
<
F
>
Query or set field or reactive energy (long time integration)
:RESET
No query form, resets all energy values
FREQuency
<
F
>
Query or set field of current or voltage signal frequency.
Depends on current or voltage synchronisation.
IMPedance:MAGnitude
<
F
>
Set the field for harmonic impedance previously selected
by the FORMat:START command.
:MAGnitude?
Query all harmonic impedances in the range specified by
the FORMat:START and FORMat:END commands.
:ANGle
<
F
>
Set the field for harmonic phase angle previously selected
by the FORMat:START command.
:ANGle?
Query all harmonic phase angles in the range specified by
the FORMat:START and FORMat:END commands.
ACQuire:RANge:VOLTage
Auto
600M Query or set voltage range
2
6
Examples:
20
ACQ:RAN:VOLT AUTO Voltage in autoranging.
60
ACQ:RAN:VOLT 200
Selects 200V range.
200
600
2000
ACQuire:RANge:CURRent
Auto
Query or set current input range
15M
1 60M
(the valid option column is fixed
50M
3 200M
by the active input, IN5, IN30,
150M 10 600M
and SHUNT)
500M 30 2
1.5
100 6
5
15
:INput
IN5
Query or set the current input or the shunt input.
IN30
SHunt
:MEAsuremo
CURRent
Query or set instrument measure mode.
VOLTage
POWer
BURSTIi
BURSTUu
FLicker
:APERture
100M
Query or set the minimal averaging time
250M
500M
Example:
1
ACQ:APER 500m Sets minimal averaging time to
500ms.
2
IEC555-2
:Hold
Run
Query or set acquisition subsystem.
Stop
Display data are held.
:QUality?
Query overload and underload of current and voltage
inputs. An integer is returned. The integer indicates the
state during the previous query
(VOLT:,CURR:,POW:,EN:, or FREQ:). For more details
refer to overload and underload register definition.