CO
2
concentration is too high
Deviation > 1%
CO
2
inlet pressure is
too high
Reduce inlet pressure
Operator
CO
2
value is fluctuating,
setpoint is not constant
Inlet pressure or flow
rate is too high
Decrease inlet pressure or flow
rate
Operator
Humidity is too low
Deviation > 10%
Water reservoir is
empty
Fill water reservoir
Operator
Sterile filter (optional) is
blocked, e.g. after dry-
running
Exchange sterile filter
Operator
Temperature is too
high
Reduce the temperature setting
Operator
Humidity is too high
Deviation > 10%
Temperature is too low Increase the temperature setting Operator
Open door to let wet air escape.
Operator
Power failure
No actions required
Unit was switched off
with activated
parameters
No actions required
7.11 Operating the tray
CAUTION!
Risk of damaging the door and the door
mechanism if the opened door is subjected to a
surface load of 26 kg or to a large localised
load.
A surface load of more than 26 kg or a large
localised load on the door can damage both the
door itself and the door mechanism.
Therefore:
– Do not place objects weighing more than 26 kg
on the door
– Do not sit on the door
– Do not use the door as a support