Transpector SPS Operating Manual
7.7.4 Electron Multiplier Replacement
NOTE:
section 7.5.3, Tools for Replacing the Electron Multiplier, on page
before continuing.
1
Remove filament lead screw (A1), loosen filament lead screw (A2), and loosen
the three screws (B) in the ion source lead connectors. Using the 1.5 x 50 mm
hex driver, unscrew the screws (C) that secure the two RF leads. Make sure to
remove the lock washers as well. There are a total of seven leads that must to
be loosened. (See
Figure 7-23 Detach leads
2
Gently pull the electrical leads away from the sensor.
CAUTION
Do not bend the leads.
B
A2
A1
C