6 - 9
Transpector MPS Operating Manual
6.5.4 Ion Source Removal
The following steps should be performed to remove the ion source:
NOTE:
section 6.5.1.2, Tools for Replacing the Ion Source, on page 6-4
before continuing.
1
While holding the clamps with the pliers, use the flat blade screwdriver to
loosen the clamps holding the filament leads (A) and ion source leads (B - one
is hidden in the figure). There are a total of six leads that need to be loosened.
See
Figure 6-6 Detach leads
2
Pull the electrical leads away from the sensor.
NOTE:
Be careful not to bend the leads.
A
B