
6 - 14
Transpector MPH Operating Manual
6
Remove the ceramic shield on the bottom of the old detector assembly.
Figure 6-12 Removing ceramic shield
7
Install the ceramic shield removed in step 6 onto the new Electron Multiplier
assembly.
8
Install the new Electron Multiplier assembly into the sensor can.
9
When installing the sensor back into the feedthrough, ensure that the two pins
on the bottom of the detector align with the two pins shown in
Figure 6-13 Pin location
10
Reverse steps 1-4 to reassemble the Transpector MPH sensor.
Detector Pins