6 - 10
Transpector MPH Operating Manual
2
Pull the electrical leads away from the sensor.
CAUTION
Do not bend the leads.
3
Using the Phillips head screwdriver, remove the gold screws (
C
- one is hidden
in the figure) that hold the ion source onto the quadrupole. There are three
screws.
Figure 6-7 Remove screws holding ion source
4
Lift the ion source off the quadrupole assembly.
5
Install the new ion source by reversing the above steps.
6
Using an ohmmeter, check:
that both filaments are not shorted to ground or to any of the lenses
that the filament leads show a filament resistance of approximately 0.8
at
the feedthrough between each filament pin and the filament common pin,
refer to
for pin out locations
C