SQM-160 Operating Manual
3.8 Display Units
The SQM-160 can display crystal measurements in several different units. To
select the display units:
1
Press Program to enter Program mode.
2
Press Prev to enter the System Menu.
3
Press Next until DSP..... is shown.
4
Turn the Control Knob left or right to select the desired display mode:
THCK — Rate in Å/s, Thickness in kÅ
nAnM — Rate in nM/s, Thickness in µM
MASS — Rate in ng/cc/s, Mass in µg (micrograms)
FREQ — Frequency in Hz
NOTE:
When Nanometers is selected a small "n" appears between the rate
and thickness displays. When Mass is selected a small "m" appears
between the rate and thickness displays.
5
Press the Control Knob to accept the choice.
6
Press Program to exit the System Menu and return to normal mode.
3.9 Crystal Life
The SQM-160 calculates the remaining crystal life based on the FMin/Max values
set in the System Menu (refer to
).
To display the remaining crystal life for the sensors used by the currently active film:
1
Press the Xtal Life switch in the front panel Control section.
2
The sensor is shown in Display 1 and the % remaining life is shown in Display 2.
3
Turn the Control Knob to display the % life of other sensors active for this film.
4
Press Xtal Life again to return to normal rate/thickness, or frequency display.
NOTE:
You cannot enter program mode while the crystal life display is active.
3.10 Zero Thickness
Before starting each film deposition, reset the SQM-160 Thickness value to zero.
To zero Thickness, press the Zero switch in the front panel Control section.
In addition to zeroing Thickness, pressing the Zero switch has these effects:
The Time display is reset to its programmed value and starts counting down.
The Thickness Setpoint and Timer relays open.
The Time SP, Thk SP, and Final Thk LEDs extinguish.
Summary of Contents for SQM-160
Page 1: ...O P E R A T I N G M A N U A L SQM 1 60 Multi Film Rate Thickness Monitor IPN 074 511 P1C ...
Page 2: ......
Page 5: ......
Page 6: ......
Page 8: ......
Page 22: ...1 10 IPN 074 511 P1C SQM 160 Operating Manual This page is intentionally blank ...
Page 30: ...2 8 IPN 074 511 P1C SQM 160 Operating Manual This page is intentionally blank ...
Page 60: ...4 12 IPN 074 511 P1C SQM 160 Operating Manual This page is intentionally blank ...
Page 76: ...5 16 IPN 074 511 P1C SQM 160 Operating Manual This page is intentionally blank ...
Page 98: ...6 22 IPN 074 511 P1C SQM 160 Operating Manual This page is intentionally blank ...
Page 110: ...8 8 IPN 074 511 P1C SQM 160 Operating Manual This page is intentionally blank ...
Page 120: ...A 10 IPN 074 511 P1C SQM 160 Operating Manual This page is intentionally blank ...