5 | Service Instructions
INFICON
56 / 76
Sensistor Sentrac-Leak Detector-ServiceManual-nip65en1-05-(1901)
5.18 Changing Probe Neck
Part
Part no.
Qty
Consumables
Hard Neck
591-963
1
LOCTITE
®
2700 or 2701
Flexible Neck
592-009
1
LOCTITE
®
2700 or 2701
Disassemble
1
Disconnect the Probe from the Sensistor Sentrac.
2
Open the Probe Cover, see Disassemble in Open Probe P60 Cover [
49].
3
Remove the Probe Sensor (A) from the Probe Neck (B) using a Sensor Key.
4
Remove the Probe PCB (D).
5
Use a heat gun to soften the adhesive in the mounting hole (C).
6
Remove the Probe Neck (B).
Assemble
1
Use Loctite 2400 on the new Probe Neck. For Flexible Neck use Loctite 2701.
2
Install the new Probe Neck. Make sure to align the hole in the hard Neck with
the mounting hole (C) in the probe.
Flex neck do not have any hole. Just tighten the neck with the threaded locking.
3
Use a Sensor Key to reinstall the Probe Sensor (A) in the new Probe Neck.
4
Reinstall the Probe PCB (D).
5
Reinstall the cables to the Probe Sensor (A).
6
Reinstall the Probe Cover, see Assemble in Open Probe P60 Cover [
Summary of Contents for Sensistor Sentrac
Page 2: ...INFICON AB Westmansgatan 47F S 58216 Lingköping Sweden ...
Page 9: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 9 76 ...
Page 10: ...3 Components INFICON 10 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 13: ...INFICON Components 3 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 13 76 ...
Page 14: ...3 Components INFICON 14 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 72: ...7 Drawings INFICON 72 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 ...
Page 73: ...INFICON Drawings 7 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 73 76 ...
Page 74: ...Index INFICON 74 76 Sensistor Sentrac Leak Detector ServiceManual nip65en1 05 1901 Index ...
Page 75: ......
Page 76: ......