RSH-600 Operating Manual
2.4 Mounting
The RSH-600 smooth sealing surface mounts to an O-ring sealed nipple (not
included). The optional adjustable flange allows the in-vacuum length to be
adjustable up to the RSH-600 original length. A flat head cover is typically used for
centered top mount applications while the angled heads may be used for either
off-center top mount or side mount applications. The customer must supply
specifications regarding mounting holes, exterior diameter of flange, and the
system mating flange.
NOTE:
The head cover must be removed from RSH-600 prior to installation.
To remove the standard stainless steel head cover: loosen the retainer ring
nut. Push the head cover inward while turning it counterclockwise until it
stops (1/16 of a turn). Pull the head cover outward.
To remove the optional copper head cover: loosen the Allen head screws
that clamp the head to RSH-600 then gently pull the head off of the
RSH-600 barrel.
The vacuum system must be equipped with a 5.08 cm (2.0 in.) diameter aperture
port. This port must have an O-ring to provide vacuum seal to the fixed flange on
RSH-600. If the head cover has been removed, simply slide RSH-600 through the
5.08 cm (2.0 in.) port. Secure RSH-600 to the port using four 1/4-24 bolts minimum.
Reinstall the head cover. See
.
Figure 2-4 Typical mounting configuration
J2
J1
15.24 cm (6 in.) Coaxial Cable
Oscillator
3 m (10 ft.) Coaxial Oscillator
Water In
Optional Flange
Water Out
Air Solenoid Valve
Cable to QCM