Inficon MAG050 Operating Manual Download Page 2

 

 

Caution

 

Vacuum component 

Dirt and damages impair the function of the 
vacuum component. 

When handling vacuum components, take ap-
propriate measures to ensure cleanliness and 
prevent damages. 

 

 

 

 

Caution

 

Dirt sensitive area 

Touching the product or parts thereof with bare 
hands increases the desorption rate. 

Always wear clean, lint-free gloves and use 
clean tools when working in this area. 

 

 

 

Precondition 

  Gauge removed from vacuum system 

 

Tools required 

  Allen wrench AF 3 

  Open-end wrench AF 7 

  Pliers for circlip 

  Polishing cloth (grain 400) or Scotch-Brite 

  Tweezers 

  Mounting tool for ignition aid 

  Cleaning alcohol 

 

 

Disassembling the Gauge 

(MAG050 

 Figure 1, MAG060 

 Figure 2) 

Precondition 

  Gauge removed from vacuum system 

Procedure 

 

Unfasten the hex head screw (3) on the magnet unit (4) 

and remove the magnet unit. 

  The magnetic force and the tendency to tilt 

make it more difficult to separate the magnet 
unit and the measuring chamber (7). 

  For reasons of tolerance, the same magnet 

unit has to be used again when reassembling 
the gauge. 

 

Remove the circlip (5) and the pole insert (6) from the 

measuring chamber (7). 

 

Loosen the 2 hex socket screws (1a) and remove the 

coaxial connector (2a). 

 

Remove the 4 (or 2) hex socket screws (8) incl. the 

lock washers (8a) on the back of the measuring cham-
ber (7). 

 

MAG050:

 Carefully remove the following items in this 

order: pressure piece (9), complete anode (10), inner 
ring (11) and FPM seal (12). 

MAG060:

 Carefully remove the following items in this 

order: pressure piece (9), washer (10b), complete 
anode (10), metal seal (11) and centering ring (12). 

The parts can now be cleaned or replaced individually 
(

 next section). 

 

 

Cleaning the Gauge 

Procedure 

DANGER

 

Adhere to the relevant regulations and take the necessary 
precautions when handling and disposing of cleaning 
agents. 

 

Cleaning the measuring chamber and the pole insert: 

  Clean the inside walls of the measuring chamber and 

the pole insert to a bright finish. Use a polishing cloth. 

Caution

 

Sealing surfaces must only be worked concentrically. 

  Rinse the measuring chamber and the pole insert with 

alcohol. 

  Dry both. 

 

Cleaning or replacing the anode (10): 

  Remove the old ignition aid (10a), for example with 

tweezers. 

  Rub the anode pin to a bright finish by means of a 

polishing cloth. 

Caution

 

Do not bend the anode. 
Do not carry out mechanical work on the ceramic 
part. 

  Rinse the anode with cleaning alcohol. 

  Dry the anode. 

  Insert the new ignition aid (10a) into the mounting tool. 

  Carefully press the anode (cleaned or new) centered 

and parallel to the tool axis into the ignition aid and in-
sert it to a depth of 

15 mm. The final position is estab-

lished only after the anode is installed. 

 
 

 

Assembling the Gauge 

Procedure 

 

MAG050:

 Insert the FPM seal (12) with the inner 

ring (11) centered into the measuring chamber (7). 
Sealing surface, seal and ceramic part must be clean 
(

 figure 1). 

MAG060:

 Insert new metal seal (11) with the centering 

ring (12) centered into the measuring chamber (7). 
Sealing surface, seal and ceramic part must be clean 
(

 figure 2). 

 

 

Carefully insert the anode (10) incl. ignition aid (10a) 

into the measuring chamber. 

 

Place the pressure piece (9) incl. Washer (10b) on the 

measuring chamber (7) and tighten the screws (8) incl. 
lock washers (8a) uniformly until the stop position is 
reached. 

 

Position the ignition aid (10a): slide the mounting tool 

over the anode pin until the mechanical stop is 
reached. 

 

Remove particles in the measuring chamber (7) by 

blowing with dry nitrogen (while the flange of the 
measuring chamber is pointing downward). 

 

Slide the pole insert (6) into the measuring chamber (7) 

up to the mechanical stop (MAG050 

 Figure 1, 

MAG060 

 Figure 2). 

 

Place the circlip (5) snugly fitting on the pole insert. 

  Visually check that the anode pin is centered 

over the hole of the pole insert (tolerated ec-
centricity 

≤0.5 mm). 

 

If possible perform a leak test 

(leak rate <10

-

9

 hPa l/s

│mbar l/s). 

 

Place the coaxial connector (2a) on the measuring 

chamber and tighten both hex socket screws (1a). 

 

Mount the magnet unit (4) and lock it with the 

screw (3). 

 

Figure 1: 

MAG050

 

 

 

 

Figure 2: 

MAG060

 

 

 
 
 
 

 

Troubleshooting 

 

Problem 

Possible cause 

Correction 

The measurement 
values indicated 
are too low 

 
Gauge contaminated 

 
Clean the gauge 

 
 

 

Spare Parts / Accessories 

When ordering spare parts, always indicate: 

  all information on the nameplate 

  description and ordering number according to spare parts 

list 

 

MAG050 

Position 

 figure 1

 

Ordering 
number 

Maintenance kit 

Inner ring 
O-ring FPM, 3.69×1.78  
O-ring FPM, 10.82×1.78 
Ignition aid 

 

(11) 

     —

      1)

 

(12) 

(10a) 

BN 846 239-T 

Repair kit 

O-ring FPM, 10.82×1.78 
Anode complete 
Inner ring 
Ignition aid 

 

(12) 
(10) 
(11) 

(10a) 

BN 846 252-T 

 

1)

  O-ring not used. 

 
 

MAG060 

Position 

 figure 2 

Ordering 
number 

Maintenance kit 

Metal seal, HNV100 9×1.6 
Centering ring 
Ignition aid 
Washer 

 

(11) 
(12) 

(10a) 
(10b) 

BN 846 241-T 

Repair kit 

Anode complete 
Washer 
Metal seal, HNV100 9×1.6 
Centering ring 
Ignition aid 

 

(10) 

(10b) 

(11) 
(12) 

(10a) 

BN 846 240-T 

 
 
 
 

 

Returning the Product 

 

 

 

Forwarding contaminated products 

Contaminated products (e.g. radioactive, toxic, 
caustic or microbiological hazard) can be detri-
mental to health and environment. 

Products returned to INFICON should preferably 
be free of harmful substances. Adhere to the 
forwarding regulations of all involved countries 
and forwarding companies and enclose a duly 
completed declaration of contamination (form 
under www.inficon.com). 

 

 

 

Products that are not clearly declared as "free of harmful 
substances" are decontaminated at the expense of the 
customer. 

Products not accompanied by a duly completed declaration 
of contamination are returned to the sender at his own ex-
pense. 
 
 
 
 

 

Disposal 

 

 

 

Contaminated parts 

Contaminated parts can be detrimental to health 
and environment. 

Before beginning to work, find out whether any 
parts are contaminated. Adhere to the relevant 
regulations and take the necessary precautions 
when handling contaminated parts. 

 

 

 

 

 

Substances detrimental to the environment 

Products or parts thereof (mechanical and elec-
tric components, operating fluids etc.) can be de-
trimental to the environment. 

Dispose of such substances in accordance with 
the relevant local regulations. 

 

 

 

Separating the components 

After disassembling the product, separate its components 
according to the following criteria: 

  Contaminated components 

Contaminated components (radioactive, toxic, caustic, or 
biological hazard etc.) must be decontaminated in accord-
ance with the relevant national regulations, separated 
according to their materials, and recycled. 

  Other components 

Such components must be separated according to their 
materials and recycled. 

 
 
 
 

 

Appendix 

 

Gas Type Dependence 

Indicated pressure (gauge calibrated for air) 

 

In the range below 10

-

5

 hPa

│mbar the pressure indication is 

linear. For gases other than air the pressure can be deter-
mined by means of a simple conversion formula: 
 

p

eff

 =  C × displayed pressure 

 

where 

Gas type 

 

Air (N

2

, O

2

, CO) 

1.0 

 

Xe 

0.4 

 

Kr 

0.5 

 

Ar 

0.8 

 

H

2

 

2.4 

 

Ne 

4.1 

 

He 

5.9 

 

These conversion factors are average values. 
 

  A mixture of gases and vapors is often involved. In 

this case accurate determination is only possible 
with a partial pressure measuring instrument, e.g. 
an INFICON quadrupole mass spectrometer. 

 
 

 

Literature 

[1]  www.inficon.com 

Operating Manual 
VGC083C 
tinb42e1 
INFICON AG, LI-0496 Balzers, Liechtenstein 

 

EU Declaration of Conformity 

 

 

We, INFICON, hereby declare that the equip-
ment mentioned below comply with the pro-
visions of the following directives: 

  2014/30/EU, OJ L 96/79, 29.3.2014 

(EMC Directive; directive relating to electro-
magnetic compatibility)

 

  2011/65/EU, OJ L 174/88, 1.7.2011 

(RoHS Directive; directive on the restriction of 
the use of certain hazardous substances in 
electrical and electronic equipment)

 

 
 

 

Products 

Cold Cathode Gauge

 

MAG050 
MAG060 
(operation with VGC083C) 

 
 

 

Standards 

Harmonized and international/national standards and specifi-
cations: 

  EN 61000-6-2:2005 

(EMC: generic immunity standard) 

  EN 61000-6-4:2007 + A1:2011 

(EMC: generic emission standard) 

  EN 61010-1:2010 

(Safety requirements for electrical equipment for measure-
ment, control and laboratory use) 

  EN 61326-1:2013; Group 1, Class A 

(EMC requirements for electrical equipment for measure-
ment, control and laboratory use) 

 
 

 

Manufacturer / Signatures 

INFICON AG, Alte Landstraße 6, LI-9496 Balzers 

15 June 2018 

15 June 2018 

 

 

Dr. Bernhard Andreaus 
Director Product Evolution 

Markus Truniger 
Product Manager 

 
 
 

 

10

9

8

7

8a

10a

11

12

3

5

1a

4

2a

7

6

10

9

8

8a

10a

11

12

7

10b

3

5

1a

4

2a

7

6

 

LI–9496 Balzers 
Liechtenstein 
Tel  +423 / 388 3111 
Fax  +423 / 388 3700 
[email protected] 

www.inficon.com 

Reviews: