Inficon MAG050 Operating Manual Download Page 1

 

Cold Cathode Gauge 

MAG050 
MAG060 

 

 

Operating Manual 

Incl. EU Declaration of Conformity 

 

tinb43e1     (2018-06) 

 
 

 

Product Identification 

In all communications with INFICON, please specify the 
information on the product nameplate. For convenient 
reference copy that information into the space provided 
below. 

 

 

 
 

 

Validity 

This document applies to products with part number 

399-840 

(MAG050, DN 25 ISO-KF) 

399-841 

(MAG050, DN 40 ISO-KF) 

399-842 

(MAG050, DN 40 CF-F) 

399-845 

(MAG060, DN 40 ISO-KF) 

399-846 

(MAG060, DN 40 CF-F) 

The part number (PN) can be taken from the product name-
plate. 

If not indicated otherwise in the legends, the illustrations in 
this document correspond to the product with vacuum 
connection DN 40 ISO-KF. They apply to the products with 
other vacuum connections by analogy. 
 
 

 

Intended Use 

The above Cold Cathode Gauges have been designed for 
vacuum measurement in the pressure range of 

MAG050: 

2×10

-9

 ... 5×10

-3

 hPa

│mbar 

MAG060: 

1×10

-10

 ... 5×10

-3

 hPa

│mbar. 

They are used together with a INFICON measurement and 
control unit of the type VGC083C. 
 
 

 

Functional Principle 

Over the whole measurement range, the measuring signal is 
output as logarithm of the pressure. 

The MAG050 / MAG060 function with a cold cathode ioniza-
tion measurement circuit (according to the inverted magne-
tron principle). 
 
 

 

Safety 

 

Symbols Used 

 

 

Information on preventing any kind of physical injury. 

 

 

Information on preventing extensive equipment and envi-
ronmental damage. 

 

 

Information on correct handling or use. Disregard can lead 
to malfunctions or minor equipment damage. 

 

  Note 

 

 

Personnel Qualification 

 

 

All work described in this document may only be carried 
out by persons who have suitable technical training and the 
necessary experience or who have been instructed by the 
end-user of the product. 

 

 

General Safety Instructions 

  Adhere to the applicable regulations and take the neces-

sary precautions for the process media used. 

  Consider possible reactions between the materials 

(

 Technical Data) and the process media. 

  Consider possible reactions (e.g. explosion) of the pro-

cess media due to the heat generated by the product. 

  Adhere to the applicable regulations and take the neces-

sary precautions for all work you are going to do and con-
sider the safety instructions in this document. 

  Before beginning to work, find out whether any vacuum 

components are contaminated. Adhere to the relevant 
regulations and take the necessary precautions when 
handling contaminated parts. 

 

 

 

DANGER

 

DANGER: magnetic fields 

Strong magnetic fields can disturb electronic 
devices like heart pacemakers or impair their 
function. 

Maintain a safety 

distance of ≥10 cm between 

the magnet and the heart pacemaker or prevent 
the influence of strong magnetic fields by anti-
magnetic shielding. 

 

 

 

Communicate the safety instructions to all other users.

 

 

 

Responsibility and Warranty 

INFICON assumes no liability and the warranty becomes null 
and void if the end-user or third parties 

  disregard the information in this document 

  use the product in a non-conforming manner 

  make any kind of interventions (modifications, alterations, 

etc.) on the product 

  use the product with accessories not listed in the corre-

sponding product documentation. 

The end-user assumes the responsibility in conjunction with 
the process media used.

 

Gauge failures due to contamination are not covered by the 
warranty. 
 
 
 

 

Technical Data 

 

Admissible temperatures 

Storage 

 

–40 °C … +80 °C 

Operation 

MAG050 

 

+5 °C … +80 °C  
(with normal cable) 

+5 °C … +150 °C  
(with high temperature cable) 

MAG060 

+5 °C … +80 °C  
(with normal cable) 

+5 °C … +250 °C  
(with high temperature cable) 

Bakeout 

MAG050 

 

+150 °C (without cable) 

MAG060 

+250 °C (without cable) 

Relative humidity 

max. 80% at temperatures 
up to +31 °C, decreasing to 
50% at +40 °C 

Use 

indoors only 
altitude up to 2000 m NN 

 

Measurement range (air, N

2

MAG050 

 

2×10

-9

 … 5×10

-3

 hPa

│mbar 

MAG060 

1×10

-10

 … 5×10

-3

 hPa

│mbar 

Accuracy (N

2

, typical) 

30% of reading 

Repeatability (typical) 

5% of reading 

Gas type dependence 

 Appendix 

 

Protection type 

IP 40 

Overpressure 

≤900 kPa (≤9 bar) 
for inert gases and tempera-
tures <55 °C only 

 

Radiation resistance 

MAG050 

 

– 

MAG060 

10

9

 rad 

 

Operating voltage 
(in measuring chamber) 

 

≤3.3 kV 

Operating current 
(in measuring chamber) 

 

≤700 µA 

 

Electrical connection 

Connector 
Type 

 

SHV 
coaxial cable 

 

Cable length between gauge 
and measurement unit 

MAG050 

 
 

max. 100 m 
(40 m if the lower limit of the 
measurement range is used 

 

 [1], Operating Manual 

VGC083C) 

MAG060 

max. 100 m 
(6 m if the lower limit of the 
measurement range is used, 

 

 [1], Operating Manual 

VGC083C) 

 

Materials on the 
vacuum side 

Vacuum connection 
Measuring chamber 
  DN 25 ISO-KF 
  DN 40 ISO-KF / CF-F 
Feedthrough isolation 
Internal seal 

MAG050 

   

MAG060 

Anode 
Ignition aid 

 
 

stainless steel (1.4306) 
 
stainless steel (1.4104) 
stainless steel (1.4306) 
ceramic (Al

2

O

3

FPM 
Ag 
Mo 
stainless steel (1.4310) 

Internal volume 

20 cm³ 

 

Dimensions [mm] 

 

 

Weight 

600 g  (DN 25 ISO-KF, DN 40 ISO-KF)  
850 g  (DN 40 CF-F) 

 
 
 
 
 

 

Installation 

 

Vacuum Connection 

 

 

D A N G E R

 

Overpressure in the vacuum system >100 kPa 
(>1 bar) 

Injury caused by released parts and harm 
caused by escaping process gases can result if 
clamps are opened while the vacuum system is 
pressurized. 

Do not open any clamps while the vacuum sys-
tem is pressurized. Use the type clamps which 
are suited to overpressure. 

 

 

 

 

D A N G E R

 

Overpressure in the vacuum system >250 kPa 
(>2.5 bar) 

KF connections with elastomer seals (e.g. 
O-rings) cannot withstand such pressures. Pro-
cess media can thus leak and possibly damage 
your health. 

Use O-rings provided with an outer centering 
ring. 

 

 

 

 

D A N G E R

 

Protective ground 

Incorrectly grounded products can be extremely 
hazardous in the event of a fault. 

The gauge must be electrically connected to the 
grounded vacuum chamber. This connection 
must conform to the requirements of a protective 
connection according to EN 61010: 

  CF connections fulfill this requirement 

  For gauges with a KF flange, use a conduc-

tive metallic clamping ring 

 

 

 

 

Caution

 

Vacuum component 

Dirt and damages impair the function of the 
vacuum component. 

When handling vacuum components, take ap-
propriate measures to ensure cleanliness and 
prevent damages. 

 

 

 

 

Caution

 

Dirt sensitive area 

Touching the product or parts thereof with bare 
hands increases the desorption rate. 

Always wear clean, lint-free gloves and use 
clean tools when working in this area. 

 

 

 

When CF vacuum connections are made, it can be advanta-
geous to temporarily remove the magnet unit (

 Removing 

the Magnet Unit). 

Mount the gauge so that no vibrations occur. Vibrations at 
the gauge cause a deviation of the measured values. 

The gauge may be mounted in any orientation. 

  To keep condensates and particles from getting into 

the measuring chamber preferably choose a hori-
zontal to upright position. 

 

Remove the protective lid and connect the product to the 
vacuum system. 

 

  Keep the protective lid. 

 

 

Removing the Magnet Unit

 

(CF vacuum connection only) 
 

  For reasons of tolerance, the same magnet unit has 

to be used again when reassembling the gauge. 

 

Tools Required 

  Allen wrench AF 1.5 

  Open-end wrench AF 7 

 

 

Procedure 

  Unfasten the hex head screw (3) on the magnet unit (4) 

and remove the magnet unit. 

  The magnetic force and the tendency to tilt 

make it more difficult to separate the magnet 
unit and the measuring chamber (7).  

  Make the vacuum connection between the gauge and 

the vacuum system. 

 

Mount the magnet unit and lock it with the hex head 

screw (3). 

 

 

Electrical Connection 

 

 

  Make sure the vacuum connection is properly 

made. 

  The VGC083C control unit must be turned off 

before any work is performed on the gauge or 
sensor cable. 

 

Connect the sensor cable to the gauge and to the INFICON 
measurement unit. 
 
 
 
 

 

Operation 

The gauge is ready for operation as soon as it has been 
connected. 

Gas type Dependence 

The measuring signal depends on the type of gas being 
measured. The value displayed is accurate for dry air, N

2

, O

2

 

and CO. It can be mathematically converted for other gases. 
This can be done by entering the corresponding calibration 
factor on the INFICON measurement unit (

 Appendix). 

 

Ignition Delay 

When cold cathode measurement systems are activated, an 
ignition delay occurs. The delay time increases at low pres-
sures and for clean, degassed gauges it is typically: 

1×10

-7

 hPa

│mbar 

 0.1 minute 

1×10

-8

 hPa

│mbar 

    1 minute 

1×10

-9

 hPa

│mbar 

    5 minutes 

1×10

-10

 hPa

│mbar 

    20 minutes  (MAG060 only) 

The ignition is a statistical process. Already a small amount 
of depositions on the inner surfaces can have a strong in-
fluence on it. 

 

Contamination 

Gauge failures due to contamination are not covered by the 
warranty. 

Cold cathode gauges are subject to contamination. The de-
gree of contamination and subsequently the accuracy of the 
measured value depend on: 

  the pressure in the vacuum chamber 

  contaminants inside the vacuum chamber (vapors, pro-

cess particles, etc.) 

  the measurement current. 

 

  To avoid extensive contamination switch the gauge 

on only at pressures of <10

-2

 hPa

│mbar. 

 

Contamination generally has the effect that the pressure indi-
cation is too low. If the contamination is severe, instability 
occurs. Contamination layers can peel off in the measuring 
chamber and cause short circuits. 

Depending on the operating conditions, cleaning may there-
fore be necessary after a few days or after a few years. 
 
 
 
 

 

Deinstallation 

 

 

 

Contaminated parts 

Contaminated parts can be detrimental to health 
and environment. 

Before beginning to work, find out whether any 
parts are contaminated. Adhere to the relevant 
regulations and take the necessary precautions 
when handling contaminated parts. 

 

 

 

 

Caution

 

Vacuum component 

Dirt and damages impair the function of the 
vacuum component. 

When handling vacuum components, take ap-
propriate measures to ensure cleanliness and 
prevent damages. 

 

 

 

 

Caution

 

Dirt sensitive area 

Touching the product or parts thereof with bare 
hands increases the desorption rate. 

Always wear clean, lint-free gloves and use 
clean tools when working in this area. 

 

 

 

Procedure 

 

Turn off the VGC083C control unit. 

 

Vent the vacuum system and disconnect the sensor 

cable from the gauge. 

 

Remove gauge from the vacuum system and install the 

protective lid. 

 
 
 
 

 

Maintenance 

Sensor failures due to contamination are not covered by the 
warranty. 
 
 
 
 

 

Cleaning the Gauge / Changing Parts 

 

 

 

Contaminated parts 

Contaminated parts can be detrimental to health 
and environment. 

Before beginning to work, find out whether any 
parts are contaminated. Adhere to the relevant 
regulations and take the necessary precautions 
when handling contaminated parts. 

 

 

 
 

Model: 
PN: 
SN:

INFICON AG, LI-9496 Balzers

ø63.5

9

2

9

7.

5

DN 40 CF-F

DN 25 ISO-KF
DN 40 ISO-KF

AF 7

Vacuum

chamber

4

3

7

 
 

 

 

Original:   English 