INFICON
Extra-long Probe | 17
074-708-P1A D-TEK Stratus OM
25 / 35
17 Extra-long Probe
D-TEK Stratus includes an extra-long replacement probe for leak checking in hard-to-
reach areas.
To install the extra-long probe:
1
Unscrew the standard probe from the D-TEK Stratus body using a 10 mm
wrench.
2
Screw on the extra-long probe to approximately 35 in·lb (4 N·m). Do not
overtighten.
3
Unscrew the probe tip from the standard probe and remove the filter (or use a
new one).
4
Insert the filter into the extra-long probe.
5
Screw the probe tip onto the extra-long probe. Do not overtighten.