22
Table 2.6 Sensor specifications
Gas type (abbreviation)
Part number
Carbon Monoxide with low Hydrogen
cross-sensitivity (CO/H
2
Low)
Hydrogen Cyanide (HCN)
17155306-G
17155306-B
Properties
Category
Toxic
Toxic
Technology
Electrochemical
Electrochemical
DualSense™ capable
No
No
Installation location
Ventis Pro4
3 or 4
3 or 4
Ventis Pro5
3 or 4
3 or 4
Operating conditions
Temperature range
a
-20 to +50 °C (-4 to +122 °F)
-30 to +40 °C (-22 to +104 °F)
RH range
a
15-95%
15-95%
Performance
Sensitivity
Measurement range
0−1000 ppm
0−30 ppm
Measurement resolution
1 ppm
0.1 ppm
Accuracy
b
Calibration gas and
concentration
100 ppm CO
10 ppm HCN
Accuracy at time and
temperature of calibration
± 5% (0-300 ppm)
± 15% (301-1000 ppm)
0 to10%
Accuracy over sensor’s
full temperature range
± 15%
± 15%
Response Time
T50
8 s
18 s
T90
12 s
65 s