18
h. Check that the ̒ 88.9cm ̕ is shown on LCD screen of the measurement unit.
* If ̒ 88.9cm ̕ is not shown on the LCD screen(error range: ±0.1cm), please contact InBody.
Reset
2) Reset: Reset all settings to factory settings.
a. Use the button to move to ‘Reset’ and then press the button.
b. Use the button to move to ‘Yes’ and then press the button.
All stored settings will be reset.
E. Maintenance
DANGER
NOTE
CAUTION
참 조
주 의
참 고
주 의
Caution
•
It may not be possible to take any measurements if the measuring tape has been damaged or dirtied.
•
It may not be possible to take any measurements if there is an excessive amount of shock placed on the
measurement unit.
•
The screw of the measurement unit along the Stand may be loosened according to environmental changes
and usage duration. Use the pressure adjusting screw on the rear of the measurement unit to adjust the
pressure appropriately.
•
Use batteries made by the same manufacturer.
•
Do not lean on the Stand or apply excessive shock to it.
•
When storing the BSM170/170B, remove the batteries and repackage.