Page 2: ...the important information below before reading this manual Failure to comply with safety warnings and regulations can cause serious injury or death DANGER NOTE 참 조 주 의 고 주 의 Failure to comply with safety cautions and regulations can cause injury or property damage DANGER NOTE 참 조 주 의 고 주 의 Warning Caution Customer Service Information InBody USA 13850 Cerritos Corporate Dr Unit C Cerritos CA 90703 ...

Page 3: ...ight Measurement A Precautionary Measurements B Measurement Instructions C Measurement Posture III Transportation and Storage A Repacking Instructions B Transportation and Storage Environment IV Frequently Asked Questions V Others A Exterior and Functions B Symbols C Specifications 4 5 5 12 18 19 20 21 22 22 23 24 26 27 User s Manual Contents ...

Page 4: ...ponents are present Please inspect each component of the BSM170 170B for damage prior to installation Footplate 1 EA Stand upper 1 EA Stand lower 1 EA Measurement unit 1 EA Measurement tape 1 EA Footplate mounting screws 4 EA Stand mounting screws 4 EA User s manual 1 EA AA batteries 4 EA Test Jig 1 EA 사용자매뉴얼 사용자매뉴얼 사용자매뉴얼 사용자매뉴얼 사용자매뉴얼 ...

Page 5: ...t the ceiling is at least 2 3 m high The maximum height of this equipment is 2 19m Keep a space of at least 3cm between the ceiling and the top of the BSM170 170B DANGER NOTE ON 참 조 주 의 참 고 주 의 Warning C Installation Instructions 1 After opening the BSM170 170B box remove the packaging materials and take all product components out Keep the packaging materials provided for repacking the equipment i...

Page 6: ...of the footplate tilt the equipment sideways towards the floor 4 Screw each of the 4 footplate mounting screws into the holes located on the opposite side of the stand insertion hole of the footplate 5 Tighten the Stand mounting screws by turning them clockwise with a Phillips screwdriver ...

Page 7: ...ed below Hold the Stand upper as vertically as possible so that it is not inserted in a slanted position DANGER NOTE UTION 참 조 주 의 참 고 주 의 Caution 8 After inserting the Stand upper into Stand lower fasten the two stands together using the mounting screws and a Phillips screwdriver Avoid using excess force to fasten the mounting screws DANGER NOTE UTION 참 조 주 의 참 고 주 의 Caution ...

Page 8: ...ement tape is completed inserted remove the sticker attached to the end of the measurement tape to affix the tape to the stand The measurement tape must be inserted with the arrow marked facing down Correct measurements cannot be taken if the measurement tape is inserted reversed DANGER NOTE UTION 참 조 주 의 참 고 주 의 Caution ...

Page 9: ...dy 10 seconds if it is connected to the InBody 1 Open the battery cover located on the right side of the LCD screen 2 Insert 4 AA batteries 3 Cover the battery socket with the battery socket lid 10 Tilt the Stand to a upright position and insert the measuring unit onto the stand as illustrated below The LCD on the measurement unit and the measurement tape must be visible on the same side DANGER NO...

Page 10: ...must move the measurement unit up and down manually By default the adjustment screw is tight to keep the measurement unit from sliding down too freely If the screw is too tight adjust the screw to the preferred tightness By default the adjustment screw is tight to keep the measurement unit from sliding down too freely If the screw is too tight adjust the screw to the preferred tightness If the scr...

Page 11: ...11 14 Turn on the BSM170 170B 15 Move the measurement unit up or down until the Move text disappears The equipment will enter the measurement standby mode and the display will change as shown below ...

Page 12: ...alue hold time When the button is pressed after a height measurement is recorded the value displayed will stay locked on the LCD screen for this amount of time even if the measurement unit is moved a Use the button to move to Hold and then press the button to select b Use the button to set a hold time for the locked height value Button mode You can lock and unlock the measurement value by pressing...

Page 13: ...or off the beep alert Beep can be set to on or off Beep setup c Save changes by pressing the button 3 Screen brightness Brightness of the LCD screen a Use the button to move to Light and then press the button b Use the button to set the screen brightness that is appropriate for the test environment Screen brightness can be set to 100 50 or off Screen brightness setup ...

Page 14: ...170B press the button to move to Exit and press the button BSM170 170B is now ready for use Offset value setup 5 Offset value Setting an offset value allows height measurements to be adjusted automatically a Use the button to move to Offset and then press the button b Move the measurement unit up or down to set an offset value appropriate for your test environment Offset value can be set up to 99 ...

Page 15: ...ter the Setup menu 2 In the Setup menu you can calibrate the equipment or reset factory default settings 1 Calibration The equipment can be calibrated to ensure correct measurement values a Raise the measurement unit to the very top of the Stand as illustrated below Right scales Left scales Calibration setup b Use the button to move to Cal and then press the button ...

Page 16: ... button to move to Yes and then press the button Calibration will begin d Slowly lower the measurement unit until you reach the bottom of the footplate and press the button The measurement unit must be lowered until it is resting against the footplate ...

Page 17: ...easurement unit so that it touches the top of the Test jig as illustrated below The Test jig must be positioned perpendicular to the footplate The Test jig must be positioned correctly f Press the button to move to Exit and then press the button BSM170 170B is now ready for use ...

Page 18: ...ion It may not be possible to take any measurements if the measuring tape has been damaged or dirtied It may not be possible to take any measurements if there is an excessive amount of shock placed on the measurement unit The screw of the measurement unit along the Stand may be loosened according to environmental changes and usage duration Use the pressure adjusting screw on the rear of the measur...

Page 19: ...the footplate adjustment screws to ensure that the equipment is level before taking measurements Use caution when lowering the measurement unit to not catch your fingers Do not run around run around or play around the footplate This may result in serious injury Children and people with limited mobility should be supervised or assisted when attempting to measure on the BSM170 170B ...

Page 20: ...ent value hold feature has been enabled in Setup pressing the button after desired measurement is displayed on the LCD locks the measurement for the set hold time Only BSM170B can connect to the InBody via its integrated Bluetooth Please refer to the InBody User s manual about Bluetooth connection guide If the BSM170B is connected to an InBody pressing the button shows the send message on the LCD ...

Page 21: ...o have accurate measurement results Let your arms rest naturally at your side Put the arms down naturally Allow your backside to gently touch the stand Touch the end of the footplate with the heels Put both the feet together Do not bend the knees Keep the legs straight ...

Page 22: ...nt tape 1EA footplate mounting screws 4EA Stand mounting screws 4EA AA batteries 4EA 4 Place 1 big polyethylene pad over the footplate 5 Close the packaging box and seal it with tape B Transportation and Storage Environment 1 The BSM170 170B should be transported or stored under the following conditions Stand lower Stand upper Footplate User s manual Test jig Accessory box Measurement unit Tempera...

Page 23: ...t tape must face downward so that when assembled the finer gradation is visible on the left hand side The correct measurement position must be maintained For details on the measurement position see C Measurement Position under II Height Measurement in this user s manual The Stand is not secure or stable Check that the bottom and joint mounting bolts are fastened tightly The measurement unit is dif...

Page 24: ...e followings are names and functions of each part of the BSM170 170B Please inspect each component of the BSM170 170B for damage prior to installation 1 Measurement unit LCD screen Shows height measurement values and equipment connection status On Off button Used for turning on off the equipment Lock button Used to lock the measured height on the measurement unit Allows for the measurement unit to...

Page 25: ... accurate measurements Stand lower Lower part of the equipment Stand 3 Lower section Stand insertion hole The Stand is inserted here for assembly with the footplate Footplate The examinee steps on the footplate when having height measured Leveling screws Consists of absolute and relative coordinates to detect the height values ...

Page 26: ...sired operation WARNING This equipment may generate or use radio frequency energy Change or modification to this equipment may cause harmful interface unless the modifications are expressly approved in the instruction manual The user could lose the authority to operate this equipment if an unauthorized change or modification is made Note This equipment has been tested and found to comply with the ...

Page 27: ... Height measurement Manual measurement Measurement unit Select between cm and inch Power DC 6V 4 AA sized 1 5V batteries External interface Bluetooth Only for BSM170B Dimension 15 35 W 19 70 L 86 23 H inch 390 W 500 6 L 2190 4 H mm Equipment Weight 8 5kg Operation Environment 10 40 50 104 30 75 RH 70 106kPa Storage Environment 10 70 4 158 10 80 RH 50 106kPa No Condensation Specifications may chang...