18
2. Replace the protective cover flap over the power input
connector of your Vesta Charger.
6.7
Charging Your OPTIMIZER Smart Mini IPG
Note:
Charging your device will take approximately 90 minutes
(if charge weekly)
Warning:
If your OPTIMIZER Smart Mini IPG is not charged
regularly, it will shut down when the battery becomes
depleted, suspending CCM therapy delivery!
Note:
Your Vesta Charger cannot be used to charge your
OPTIMIZER Smart Mini IPG until the AC Adapter is disconnected
from your Vesta Charger.
Caution:
Your Vesta Charger should not be operated close to
other electronic equipment. If sufficient spatial separation cannot
be maintained, your Vesta Charger needs to be monitored to
ensure normal function.
Warning:
The Vesta Charger must not be used onboard an
aircraft.
Warning:
When onboard a ship, request permission from the
ship’s crew prior to using your Vesta Charger.
To charge the battery of your OPTIMIZER Smart Mini IPG,
perform the following steps:
1. Assume a stationary, comfortable sitting position, ideally
reclining at a 45° angle (such as on a sofa or armchair).
2. Determine the location of your OPTIMIZER Smart Mini
IPG (typically right upper chest area). Drape the wand
cable loosely around your neck and then place the flat
side of the Vesta charging wand (the side with the four
blue rubber screw covers) directly over your OPTIMIZER
Smart Mini IPG implant site (over your clothes). To
prevent the charging wand from becoming displaced while
charging your implanted OPTIMIZER Smart Mini IPG, you
may attach the charging wand cable clip to your clothing.
3. Start the charging process by pressing the
Power
Button
, holding the button down for 1-2 seconds, and
then releasing it.
See Figure 12
.