Impex LD8000 User Manual Download Page 11

Laser Diode Driver LD8000 

11 

Ver. 1.0 

when  using  the  propriate  laser.  The  button 

Defaults

  loads  default  values, that  are  programmed 

into the device by manufacturer. 
The 

Copyright

 notice and the 

Impex HighTech

 logo are linked to the Impex HighTech website. 

 

 

In  this  row,  there  are  value  indicators  for  quick  control  and  monitoring  switches  (green  round 
buttons). Only if the monitoring switch is enabled (i.e. it is lighting lime-green), the appropriate 
actual  value  is  transferred  from  the  device  and  monitored  in  the  field  to  the  left  and  on  the 
monitor page. The rectangular indicators to the right light red, if something is wrong (from left to 
right): laser diode current, laser diode voltage, out of temperature range (upper or lower limit). 

System card 

 

Figure 8. System card 

COM Port Number 

sets the number of the virtual COM Port, that is given by the USB to serial 

converter driver (see chapter Installing USB driver).

 

Baud Rate 

sets the transfer speed and should be set to 57600 bps in current revision.

 

The  window 

Communication  Status 

shows  the  error  messages  if  some  communication  error 

occur. The 

Error

 indicator lights up in such case.

 

The 

Clear Error 

button clears the Communication Status window and the Error indicator turns 

off.

 

If the device is turned off or USB cable was removed, the connection is lost. After the problem is 
solved  and  the  driver  is  connected  properly,  press  the 

Reconnect

  button  to  reactivate 

communication.

 

The  switch 

Manual  Control

  enables  or  disables  keys  and  the  knob  at  the  front  panel  of  the 

device. This function is used to prohibit a change of values if the laser is operated remotely.

 

The  fields 

Laser  Diode  Life  Time

Driver  Serial  Number

  and 

Driver  Software  Version

  are 

self-explanatory.

 

Summary of Contents for LD8000

Page 1: ...Version 1 0 20 November 2008 IMPEX HighTech GmbH Hovesaatstr 6 48432 Rheine Germany Phone 49 0 5971 981 650 FAX 49 0 5971 981 659 Laser Diode Driver LD8000 User Manual...

Page 2: ...ertification of any kind Before connecting or disconnecting any cable switch off the device main power switch Meet all necessary precausions to except damages throuh ESD electrostatic discharge Always...

Page 3: ...t 8 Indicators 8 4 PRINCIPLES OF OPERATION 9 4 1 Standalone operation 9 Navigating through the menu 9 Rearranging the list 9 Changing a parameter or a value 9 4 2 The LD Control software 10 Installing...

Page 4: ...generating photo diode input for constant power mode analog modulation input for external feedback circuit User friendly control panel allowing quick access to common parameters and values USB control...

Page 5: ...2 5 Pulse overshoot 1 2 TTL Low level V 0 0 8 TTL High level V 3 5 5 Temperature setting range C 10 60 Temperature stability C 0 01 0 02 TEC controller voltage V 11 5 11 5 TEC controller current A 8...

Page 6: ...e input BNC 3 Up button lit if operable 11 Laser head connector HD D SUB25 4 OK button lit if operable 12 Interlock connector 3 5mm TS connector 5 Down button lit if operable 13 Power supply connector...

Page 7: ...Sensor 21 12V 4 Crystal TEC anode 13 Crystal NTC Sensor 22 Ground 5 Harm TEC cathode1 14 Harm NTC Sensor1 23 Photo diode signal input 6 Harm TEC anode1 15 Harm NTC Sensor1 24 Laser diode cathode 7 Las...

Page 8: ...meter or scope can be connected to this connector to monitor the power Modulation input In external analog modulation mode voltage at the inner pin of this connector is converted to diode current with...

Page 9: ...character of the line starts to blink and move the parameter with the Up and Down buttons When the desired position is achieved store the state with the OK button If you don t store the state the prev...

Page 10: ...quipped with online description Get the mouse pointer over an element for view seconds and the description will pop up The application is divided into five parts represented by a card index On the Sys...

Page 11: ...em card COM Port Number sets the number of the virtual COM Port that is given by the USB to serial converter driver see chapter Installing USB driver Baud Rate sets the transfer speed and should be se...

Page 12: ...LD Voltage Error indicator green rectangular lights up The Gate Mode is available in modulation modes with internal pulse generation See chapter 4 3 Laser modulation modes for details Constant output...

Page 13: ...you have to specify the Upper and Lower Limits Within this window the laser operation is ensured When the temperature exceeds this limits the laser and the corresponding temperature controller are tu...

Page 14: ...ges use the one on the right side of the graph by default At the upper right corner you can set up display options such as line style and color for each value individually Just click on the line picto...

Page 15: ...urrent high level means Target current 50 Periodic Burst this mode is similar to the internal pulse mode but the given Number of pulses is followed by given number of Suppressed pulses During suppress...

Page 16: ...Laser Invert Gate Signal bool RLPCA Photo Curr Laser Photo Current Actual uint16 uA RLPCT Photo Curr Laser Photo Current Target uint16 uA RLPCC Ph Cur Contr Laser Photo Current Control bool RLMBN Burs...

Page 17: ...tual float C RLTC LD Temp Ctrl Laser Temp Controller Enable bool RCTT Crystal Temp Crystal Temperature Target float C RCTA Crystal Temp Crystal Temperature Actual float C RCTC Cryst T Ctrl Crystal Tem...

Page 18: ...t properly packed it will be damaged in shipping and the warranty will be avoided This warranty shall not apply to any defect failure or damage caused by improper use failure to observe proper operati...

Reviews: