22
Operation
The charger starts charging automatically as soon as a battery pack is installed and the charger is
plugged in. Usually, the battery packs are brought into contact by the plugs which are enclosed the
supply. If the red LED after the test phase (about 10 seconds) still keeps flashing, check the polarity of
the battery pack. Please gather the changing of the polarity (+/-) of the battery pack out of illustration
No. 5. If the red LED is flashing again after the test phase and does not switch over to permanent light
for charging, either the battery pack is defect or the battery pack might also not have the right number
of cells (less than 4 cells (4,8V) or more than 10 cells (12,0 V)). The test phase is followed by the
charging procedure (red LED is permanent on). After the charging procedure, the charger switches
automatically over to trickle charge (green LED is flashing; red LED is off). The starting of the
discharging procedure occurs by pressing ( about 2 seconds) the button for discharging (3). However,
this discharging procedure should only take place after the test phase. After discharging, which can in
individual cases last for several hours, the charger automatically switches over to charging.
Use in motor vehicles and trucks ( only ACS 410 mobile)
The ACS 410 mobile can be plugged into a 12-32 V DC outlet. If you use it on DC you have to connect
the provided DC cable to the cigarette lighter.
Note!! In passenger cars (12-16 V DC) only 4 to 6 cells battery packs (4.8-7.2 V) can be charged. In
trucks (24-32 V DC) you can charge 4 to 10 cells battery packs (4.8-12 V).
Environmental reference
Rechargeable batteries are not to be disposed in domestic waste. Surrender used batteries to your
dealer or rather to the battery collecting point