Basic Alignment with the Calibration Set
5.1
What is a basic alignment?
The basic alignment serves to compensate the cable length and tolerances of the probe
mechanics (thickness of the rod coating, rod length, etc.). After two measurements, one in dry
and one in water saturated glass beads, the calibration data is calculated and stored in the
TRIME probe’s connector.
Every TRIME probe must be calibrated before it can supply proper measurement results. It is
very important that the probe is calibrated with the device it belongs to. This calibration is
regularly done by IMKO before the delivery of new probes.
Please note that probes can only be used with the TRIME device they were calibrated with!
5.1.1
What are the benefits of the calibration set for the user?
With the calibration set you can easily calibrate probes on your TRIME instrument by yourself.
The calibration set offers the following advantages to you:
1. If defective probe rods must be changed you are able to make the required basic
alignment by yourself.
2. You can calibrate probes purchased later with your TRIME-instrument by yourself.
3. If you have two or more instruments you can change the configuration of instruments
and probes anytime by a new calibration of the appropriate probes to the specific
instruments.
You would have to send instruments and probes to us in any of the above cases where basic
alignment is required if you have no calibration set. IMKO will charge calibration costs. Moreover
you will save the forwarding charges and you will not have to interrupt your measuring campaign,
if you calibrate by yourself.
However, the calibration set allows only to make a basic alignment of the TRIME probes. You
can not change the standard calibration for specific materials with it. For this purpose a
measurement data set must be created with this specific material. In order to calculate the
calibration data for this data set and to download it into the TRIME-probe you need the
calibration program SMCAL.
Coax a
c ble
Calibration
Connector
Probe
connector
TRIME
Probe
dry glass
beads
water
saturated
glass beads
Start