Page 32
Step-by-Step Procedure
Step 1:
Clean the tip with soapy water, ensuring that the
mirror is clean and free of smudges, stains, or any residue.
Avoid using abrasive cloth materials as this will scratch
the mirror.
Step 2
: After drying the tip exterior, carefully dry the
interior and mirror with lint free wipes ensuring you do
not scratch the sur- face. The mirror should be free of any
noticeable debris or water spots.
Step 3:
Insert and seal the tip into a sterilization pouch.
Make sure the seal is airtight. Each tip should be
packaged individually.
Step 4:
Sterilize the wrapped tip in a steam autoclave at
following parameters:
•
132°C (270°F) at 4 minutes, or
•
134°C (273°F) at 4 minutes
Step 5:
Ensure the dry cycle is complete prior to
removing the tip from the autoclave. If the pouch is damp
with moisture, pro- per sterilization cannot be
guaranteed.
WARNING
: Always autoclave the tip wrapped in a sealed sterilization
pouch; failure to do so will result in permanent stains on the mirror.
WARNING
: Do not autoclave the handpiece of the device.
NOTE
: Tips should not be placed in an ultrasonic cleaner or any cold
sterile solutions. The sterilant solutions will leave a sticky residue or
film on the mirror when drying.
WARNING
: Do not remove the pouch before the sterilizer completes
its full dry cycle. If the pouch is wet or has any signs of moisture, this
can potentially leave water spots on the mirror which can affect image
quality during scanning.
WARNING
: Use extreme caution when cleaning the mirror as it is
very delicate and is prone to scratching.