![ILX Lightwave LDX-32420 User Manual Download Page 86](http://html1.mh-extra.com/html/ilx-lightwave/ldx-32420/ldx-32420_user-manual_2044776086.webp)
FU N C T I O N S A N D FE A T U R E S
Using the LDX-32420 Current Source’s Trigger Function
01_10
LDX-32420
69
C H A P T E R
5
Using the LDX-32420 Current Source’s Trigger Function
For applications where you need to synchronously initiate a measurement task
from a remote instrument with the LDX-32420 Current Source, the Current Source
offers a trigger output signal. The TTL pulse is initiated with any remote change in
setpoint of the laser current source. A typical application for utilization of this
feature is laser characterization tasks where an L-I curve is generated. For a
programmed step in laser current, a measurement can be triggered for each step
in laser current. The trigger output is enabled automatically in remote mode.
The trigger output is available via a standard BNC connector on the rear panel of
the LDX-32420 Current Source. See Figure 5.1. The TTL pulse is approximately
13
s wide and has a delay time of about 10-12 ms after the current setpoint is
changed via remote command. This delay does not include transmission time.
The time for any command to be sent depends on the interface hardware and
software control. The current setpoint may be changed by the LAS:LDI, LAS:INC,
or LAS:DEC commands.
Note:
The minimum step change time required for a corresponding trigger pulse is 20 ms.
Care should be used with respect to the timing of any setpoint commands in relation to the
actual hardware function. The INC and DEC commands have a minimum step time of
about 20 ms that allows for the hardware initiation of a setpoint change and the settling of
the output. The LAS:LDI command does not have any fixed delay for settling. With this
command, some trigger pulses may be missed if the program step time is less than the
20 ms minimum.
A one shot trigger pulse will occur on power up of the instrument due to the states of the
processor I/O.
Figure 5.1
LDX-32420 Rear Panel
Summary of Contents for LDX-32420
Page 3: ......
Page 9: ...L I S T O F FIG U R E S vi LDX 32420...
Page 11: ...L I S T O F TA B L E S viii LDX 32420...
Page 17: ...WA R R A N T Y xiv LDX 32420...
Page 41: ...O P E R A T I O N Operating a Laser in Constant Power P Mode 24 LDX 32420 C H A P T E R 2...
Page 57: ...RE M O T E O PE R A T I ON Error Messages 40 LDX 32420 C H A P T E R 3...
Page 103: ...C A L IB R A TI O N A N D TR OU BL E S HO O T IN G Error Messages 86 LDX 32420 C H A P T E R 6...