C A L I B R A T I O N A N D T R O U B L E S H O O T I N G
Troubleshooting Guide
01_10
LDX-32420
83
C H A P T E R
6
Symptom
Corrective Actions
Unable to adjust output
Check the I LIMIT parameter for the output range in use, see
that it is set above 0 A.
Power mode operation has high
output current, but little or no
power measured
Check back panel PD BIAS; if set too low, may act as an open
feedback loop; if in doubt, set the PD BIAS to mid range (2.5 V)
Output exceeds power limit
The “power limit” is not a hardware limit. It only serves as a
warning that the power measurement has exceeded the limit
setpoint. Normally this limit will cause the output to shut off. This
feature can be defeated via the LAS:ENABLE:OUTOFF
command.
Open circuit error occurs during
calibration
Check load connections; check that measuring meter does not
auto-range (use non-auto-ranging modes).
Calibration is aborted
unintentionally
Calibration modes will be aborted if an open circuit is detected.
VOLTAGE LIMIT indicator blinks
This indicates a voltage limit error; check laser connections. A
high impedance may cause this condition.
Open circuit error E503 or voltage
limit error E505 prevents output
from reaching desired value
The LDX-32420 instruments have an adjustable laser
compliance voltage. Check to see if the LASER voltage limit
setting is too low (Chapter 2); check laser connections.
Summary of Contents for LDX-32420
Page 3: ......
Page 9: ...L I S T O F FIG U R E S vi LDX 32420...
Page 11: ...L I S T O F TA B L E S viii LDX 32420...
Page 17: ...WA R R A N T Y xiv LDX 32420...
Page 41: ...O P E R A T I O N Operating a Laser in Constant Power P Mode 24 LDX 32420 C H A P T E R 2...
Page 57: ...RE M O T E O PE R A T I ON Error Messages 40 LDX 32420 C H A P T E R 3...
Page 103: ...C A L IB R A TI O N A N D TR OU BL E S HO O T IN G Error Messages 86 LDX 32420 C H A P T E R 6...