R E M O T E O P E R A T I O N
LDX-32420 Current Source Command Set
01_10
LDX-32420
29
C H A P T E R
3
The following are examples of invalid syntax. These command strings would
produce an erroneous result, as explained:
For further clarity in programming, the (Boolean) values of one (1) and zero (0)
may be used or their names as indicated in the following table:
Table 3.1
Substitute Parameter Names
If multiple parameters are expected, they should be separated with commas:
LAS:INC 100,50
A query has no space between the mnemonic and the question mark:
LAS:LDI?
When sending commands through the GPIB bus, the LDX-32420 uses a
terminator of <NL><^END> (new line with EOI). For users whose GPIB driver
defaults expect a carriage return in the terminator, <CR><NL><^END>, the
TERM
command may be used for convenience (see
TERM
command in Chapter 4). For
more information, refer to your GPIB driver configuration manual.
Commands sent through the USB bus must always be terminated with new line,
<n1>.
LAS:MODE MDP
Missing colon, MODE? expected
LAS:MODE:ILBW DEC
Missing semicolon, DEC command generates an error
LAS:DIS ?
Space not allowed before question mark, DIS command
expected
Las:LDI33;dis?
Space missing between LDI command and the parameter
value, 33
Substitute Name
Value
ON
1
OFF
0
OLD
1
NEW
0
TRUE
1
FALSE
0
SET
1
RESET
0
Summary of Contents for LDX-32420
Page 3: ......
Page 9: ...L I S T O F FIG U R E S vi LDX 32420...
Page 11: ...L I S T O F TA B L E S viii LDX 32420...
Page 17: ...WA R R A N T Y xiv LDX 32420...
Page 41: ...O P E R A T I O N Operating a Laser in Constant Power P Mode 24 LDX 32420 C H A P T E R 2...
Page 57: ...RE M O T E O PE R A T I ON Error Messages 40 LDX 32420 C H A P T E R 3...
Page 103: ...C A L IB R A TI O N A N D TR OU BL E S HO O T IN G Error Messages 86 LDX 32420 C H A P T E R 6...