Figure 26
Wash Cartridge
Prepare for a Manual Post-Run Wash
1
Combine the following volumes to result in 0.12% NaOCl.
u
5% NaOCl (31 µl)
u
Laboratory-grade water (1269 µl)
2
Add 1.3 ml of 0.12% NaOCl to the wash cartridge.
The correct reservoir is equivalent to position #31 on the prefilled reagent cartridge.
Figure 27
Positions for NaOCl and Wash Solution
A
0.12% NaOCl
B
Wash solution
3
Combine the following volumes to result in a 0.05% Tween 20 wash solution.
u
100% Tween 20 (40 µl)
u
Laboratory-grade water (80 ml)
4
Add 80 ml wash solution to the wash cartridge.
The correct reservoir is equivalent to position #40 on the prefilled reagent cartridge.
5
From the Home screen, select Perform Wash, and then select Manual post-run wash.
Document # 1000000002695 v02 Material # 20014309
For Research Use Only. Not for use in diagnostic procedures.
33
MiniSeq System Guide