NOTE
After you select Home, it is not possible to return to view run metrics. However, run metrics are accessible on
BaseSpace, from a networked computer using Sequencing Analysis Viewer, or from a networked computer
using Local Run Manager.
Cycles for Run Metrics
Run metrics appear at different points in a run.
u
During the cluster generation steps, no metrics appear.
u
The first 5 cycles are reserved for template generation.
u
At cycle 6, raw cluster density and cycle 1 intensities are available.
u
After cycle 25, clusters passing filter, yield, and quality scores are available.
Sequencing Analysis Viewer
The Sequencing Analysis Viewer software shows sequencing metrics generated during the run. Metrics
appear in the form of plots, graphs, and tables based on data generated by RTA and written to InterOp files.
Metrics are updated as the run progresses. Select Refresh at any time during the run to view updated
metrics. For more information, see the
Sequencing Analysis Viewer User Guide (part # 15020619)
.
The Sequencing Analysis Viewer is included in the software installed on the instrument computer. You can
also install Sequencing Analysis Viewer on another computer linked to the same network as the instrument to
monitor run metrics remotely.
Data Transfer Status
Depending on the analysis configuration selected, an icon appears on the screen during the run to indicate
connection status.
Status
BaseSpace
BaseSpace Onsite
Local Run Manager
Standalone
Connected
Transferring data
Disconnected
Disabled
It is possible that multiple icons appear on the screen. For example, if run data are being transferred to
BaseSpace and an additional output folder location, a BaseSpace icon and a standalone icon appear.
Run Copy Service
The MiniSeq System Software Suite includes a Run Copy Service. As RTA generates files, the service copies
files to the specified output folder location.