Load the Manifold
From the Manifold screen, load the manifold from the same cluster kit as the flow cell.
1
Inspect the sippers on the sipper comb for damage. Make sure that the black rubber
gaskets are evenly seated.
2
Position the manifold over the flow cell with the sipper comb pointing toward the front
of the cBot.
3
Align the manifold with the guide pins on the thermal stage and seat the manifold
over the flow cell. Seat evenly to form a tight seal.
4
Select the
Manifold seated over flow cell
checkbox.
5
Close the flow cell clamp to secure the manifold.
Figure 13
Close the Flow Cell Clamp
6
Select the
Flow cell clamp closed
checkbox.
7
Connect the outlet end of the manifold to the outlet port in the wash reservoir. Make
sure that the outlet is evenly seated.
Figure 14
Secure Outlet End
A
Outlet Port
B
Output Clamp
8
Snap the outlet clamp closed to secure the outlet end of the manifold.
9
Select the
Outlet clamp closed
checkbox.
10
Align the sipper comb with the 2 metal guide pins on the front of the thermal stage.