CAUTION!

  

 

 

5.3   Electrical Connections 
The Centurion is provided with a 24VDC AC adapter. This AC adapter provides 
electrical ground to the unit via a direct connection through the adapter to the wall 
outlet. For this reason the unit should be used with this AC adapter only. The unit’s line 
cord should be plugged in to a standard 3-terminal grounded receptacle with line 
voltage and frequency of 100-240 VAC, 50/60 Hz. Note that an instrumentation ground 
is provided on the rear of the unit for testing and calibration purposes.  
 

 
 

SECTION 6  Operation 

 

The Centurion’s ON/OFF switch is located at the rear of the unit. The unit should be 
turned on using the switch when 24VDC is provided via the AC adapter. Do not ‘hot 
plug’ the AC adapter into a unit when the switch is in the ON position. Switch the unit 
off, then insert the 24VDC connection into the unit. 
 
Fan speed is adjustable at the unit’s rear panel using a potentiometer adjustment tool or 
small screwdriver. Turning the control clockwise increases the fan output; turning 
counter clockwise reduces the airflow. Note that the time required to neutralize static is 
dependent on the fan speed setting. For fastest charge neutralizing performance, the 
unit should be adjusted for the highest fan speed. 
 
The Ion Balance adjustment located on the rear of the unit tunes the Centurion’s offset 
voltage to zero. Typically this adjustment is made as part of scheduled maintenance or 
setup. Adjustments to this control should not be made without using the correct 
calibration instrumentation. 
 

 

 

SECTION 7  Maintenance

 

 
7.1 Ion Balance Calibration 
The Centurion is delivered from the factory calibrated to a zero volt offset set point. As 
part of regular maintenance, or if the unit’s fan speed is changed, the ion balance 
calibration should be checked.  For calibration, a charge plate monitor (CPM) such as 
SIMCO’S EA-4 is required. The CPM should have its ground connected to the 
instrument ground provided on the rear panel of the Centurion; this jack accepts a 
standard banana plug. For most accurate calibration results, the unit should be 
operated at least 30 minutes and the test location should not be subject to other 
ionization sources. The operator should wear a grounding wrist strap. With the charge 
plate located 12” from the outlet grille of the unit, the ion balance control can be 
adjusted for 0V using a potentiometer adjustment tool or small screwdriver.  
 
 

 

Summary of Contents for SIMCO CENTURION

Page 1: ...single fan DC ionizer offers superior balance performance with its novel emitter array design and high gain dual feedback circuitry The Centurion eliminates static charges cleanly quickly and reliably...

Page 2: ...made to the output of the emitter array are seen promptly as ion output changes in the air The resulting balanced stream of ionized air is delivered to the work area to provide static neutralization...

Page 3: ...an Speed Adjusted using a potentiometer adjustment tool Adjustment located on the rear panel of the unit Offset Balance Adjustable to zero using a potentiometer adjustment tool Adjustment access is lo...

Page 4: ...e unit in flammable volatile or explosive atmosphere 6 Do not insert objects through the unit s intake or outlet grilles Damage to the ionizer and or personal injury may result SECTION 5 Installation...

Page 5: ...ize static is dependent on the fan speed setting For fastest charge neutralizing performance the unit should be adjusted for the highest fan speed The Ion Balance adjustment located on the rear of the...

Page 6: ...removal from its pivot pins The emitters are installed in sockets for easy insertion and removal Using a long nose pliers pull the emitters out along the radial direction gripping the shaft only takin...

Page 7: ...n no event shall SIMCO be liable for incidental or consequential damages except where state laws override This warranty extends to the original purchaser and is not transferable No person agent distri...

Reviews: