IKA ULTRA-TURRAX Tube Drive P control Manual Download Page 7

21

Ö

 Then, the system check screen will be displayed for several seconds.

Symbols are shown on the screen, including the working mode set-
ting, the target speed, the maximum rotation speed (rpm) and timer 
maximum value etc. of the device.

Factory setting

The device is supplied with following factory setting:

Menu Settings:
mode  

A

max rpm   

6000 

rpm

timer max   

30 

minute

language  

English

Working Setting:
target rpm   

400 

rpm

timer sec   

off

rev sec 

 

off

10

Resetting parameters to factory setting

10.1

To reset the parameters to the factory settings, proceed as following:

• Pull out main plug from the device.
• Hold the controller buttons “Menu/Back” and “Timer” simultane-

ously and re-insert main plug in power outlet.

• The unit is now reset to the factory settings.

c

g

d

e

f

h

Symbols Explanation

c

 Maximum rotation speed of the device (If “Turbo” function is 

activated, the maximum rotation speed could reach 8000 rpm).

d

 Rotation symbol (The symbol will change into 

3

 or 

4

 if  “Rev” 

function is activated).

e

 Rotation speed setting.

f

 Operation mode.

g

 “Timer” maximum value of the device.

h

 “Timer” symbol.

Ö

 After the system check screen, the working screen will appear 

and the device is ready for operation.

Note:

 

Figure above shows specified values corresponding to the 

default settings at delivery.

Summary of Contents for ULTRA-TURRAX Tube Drive P control

Page 1: ...IKA ULTRA TURRAX Tube Drive P control 20000005139 UTTD P control_042018...

Page 2: ...ommissioning 20 9 Switching on 20 10 Factory setting 21 10 1 Resetting parameters to factory setting RESET 21 11 Set menu parameters 22 11 1 Operation mode 22 11 2 Operation language 22 11 3 Maximum r...

Page 3: ...the drive unit prior to operating the device The tube must only be attached and removed while the motor is stationary The IKA tubes must always be closed when the device is in operation Switch off the...

Page 4: ...T 50 Dispersing Stirring 5 15 ml 15 50 ml A B P1 P9 Turbo Rev with Tube BMT 20 with Tube BMT 50 Crushing Mixing wet dry 2 15 ml 15 50 ml A B P1 P9 Turbo Rev with Tube DIS 50 with Tube DIS 300 X Crushi...

Page 5: ...as described in RS 232 interface Command syntax and format The following applies to the command set Commands are generally sent from the computer Master to the device Slave The device sends only at t...

Page 6: ...on is pressed after this command is sent to device Remote symbol is displayed RESET Switch to normal operation Device Keypad available again and clear the Re mote symbol IN_PV_X X 4 Read actual speed...

Page 7: ...ings proceed as following Pull out main plug from the device Hold the controller buttons Menu Back and Timer simultane ously and re insert main plug in power outlet The unit is now reset to the factor...

Page 8: ...the Rotating Pressing knob B to select the mode item in the main menu 11 Operation mode 11 1 Press the Rotating Pressing knob B Turn the Rotating Pressing knob B to select the mode item in mode P1 an...

Page 9: ...nge from 10 seconds to 30 minutes This is achieved in magnitudes of 10 seconds In remote control mode via computer adjusting magnitude is 1 second The default setting max rpm is 6000 rpm in the device...

Page 10: ...v value with the Rotating Pressing knob B the value can be set from 10 to 60 seconds where adjustment magni tude is 1 second Note If both Rev function and Timer function are activated the Rev value sh...

Page 11: ...designation of the spare part see www ika com spare parts diagram and spare parts list Turbo function will be stopped and rotation speed will return to original setting A symbol appears to indicate th...

Page 12: ...lts resulting from improper use or insufficient care and mainte nance contrary to the instructions in this operating manual See more accessories on www ika com Repair Please send in device for repair...

Page 13: ...A Mode B Mode P1 P9 rpm 400 6000 in nitely adjustable adjustment magnitude 10 rpm up to 8000 up to 6000 up to 6000 up to 6000 Speed setting rotating pressing knob Speed display digital Timer 10 secon...

Reviews: