IKA ULTRA-TURRAX Tube Drive P control Manual Download Page 10

24

Setting working parameters

12

Set “target rpm” within “rpm max” range in the working screen 
by rotating the “Rotating/Pressing” knob (B). Adjustment magni-
tude is 10 rpm. In remote control mode via computer, adjusting 
magnitude is 1 rpm. 

• 

Setting “target rpm”  in working screen

)

Turn the “Rotating/Pressing” knob (B) to change the value.

)

Press “Rotating/Pressing” knob (B) to start working.

Ö

Rotation symbol ( ,

3

,

4

) indicates running.

Target speed (target rpm) 

12.1

Timer function

12.2

Reverse (Rev) function

12.3

Turbo Function

12.4

The rotation speed can increase to 8000 rpm in several seconds by 
pressing down „Turbo“ button (F) during operating. For tube protec-
tion, Turbo function can only last 60 seconds at most if you press 
down „Turbo“ button (F).  Several seconds later, Turbo function can 
be activated again after symbol has disappeared.

• 

Activate and hold “Turbo” function in working screen:

)

 Press down „Turbo“ button (F) over 2 seconds when the de-

vice is working, to activate Turbo function.

Ö

 The Turbo mark 

 appears and changes alternately to in-

dicate that the Turbo function is activated and rotation speed 
will increase to 8000 rpm in a short time.

)

 Release “Turbo” button (F).

When setting “Rev” value with the “Rotating/Pressing” knob (B), the 
value can be set from 10 to 60 seconds where adjustment magni-
tude is 1 second.

Note:

 If both “Rev” function and “Timer” function are activated, 

the “Rev” value should be less than “Timer” value. Otherwise, the 
reverse function will not work.
If the reverse function is activated, press “Rev” button (E) dur-
ing running, the reverse function will be deactivated. Press “Rev” 
button (E) again, the reverse function will be reactivated and the 
screen will indicate “Rev” value setting.

• 

Setting “Rev” timer (rev sec) in working screen

)

 Press “Rev” button (E) to activate the reverse setting in the 

working screen.

)

 Turn the “Rotating/Pressing” knob (B) to change the value.

)

 Press “Rev” button (E) again to complete the setting.

Ö

 Current values are stored.

Set “Timer” within “timer max” range in the working screen by ro-
tating the “Rotating/Pressing” knob (B). Adjustment magnitude is 10 
seconds. In remote control mode via computer, adjusting magnitude 
is 1 second.
If the “Timer” is deactivated when the device is started, the device 
will automatically count from 0 to 30 minutes and then the unit will 
stop running.

• 

Setting “Timer” value (timer sec) in working screen

)

 Press “Timer” button (D) to activate the timer setting in the 

working screen.

)

 Turn the “Rotating/Pressing” knob (B) to change the value.

)

 Press “Timer” button (D) again to complete the setting.

Ö

 Current values are stored.

Summary of Contents for ULTRA-TURRAX Tube Drive P control

Page 1: ...IKA ULTRA TURRAX Tube Drive P control 20000005139 UTTD P control_042018...

Page 2: ...ommissioning 20 9 Switching on 20 10 Factory setting 21 10 1 Resetting parameters to factory setting RESET 21 11 Set menu parameters 22 11 1 Operation mode 22 11 2 Operation language 22 11 3 Maximum r...

Page 3: ...the drive unit prior to operating the device The tube must only be attached and removed while the motor is stationary The IKA tubes must always be closed when the device is in operation Switch off the...

Page 4: ...T 50 Dispersing Stirring 5 15 ml 15 50 ml A B P1 P9 Turbo Rev with Tube BMT 20 with Tube BMT 50 Crushing Mixing wet dry 2 15 ml 15 50 ml A B P1 P9 Turbo Rev with Tube DIS 50 with Tube DIS 300 X Crushi...

Page 5: ...as described in RS 232 interface Command syntax and format The following applies to the command set Commands are generally sent from the computer Master to the device Slave The device sends only at t...

Page 6: ...on is pressed after this command is sent to device Remote symbol is displayed RESET Switch to normal operation Device Keypad available again and clear the Re mote symbol IN_PV_X X 4 Read actual speed...

Page 7: ...ings proceed as following Pull out main plug from the device Hold the controller buttons Menu Back and Timer simultane ously and re insert main plug in power outlet The unit is now reset to the factor...

Page 8: ...the Rotating Pressing knob B to select the mode item in the main menu 11 Operation mode 11 1 Press the Rotating Pressing knob B Turn the Rotating Pressing knob B to select the mode item in mode P1 an...

Page 9: ...nge from 10 seconds to 30 minutes This is achieved in magnitudes of 10 seconds In remote control mode via computer adjusting magnitude is 1 second The default setting max rpm is 6000 rpm in the device...

Page 10: ...v value with the Rotating Pressing knob B the value can be set from 10 to 60 seconds where adjustment magni tude is 1 second Note If both Rev function and Timer function are activated the Rev value sh...

Page 11: ...designation of the spare part see www ika com spare parts diagram and spare parts list Turbo function will be stopped and rotation speed will return to original setting A symbol appears to indicate th...

Page 12: ...lts resulting from improper use or insufficient care and mainte nance contrary to the instructions in this operating manual See more accessories on www ika com Repair Please send in device for repair...

Page 13: ...A Mode B Mode P1 P9 rpm 400 6000 in nitely adjustable adjustment magnitude 10 rpm up to 8000 up to 6000 up to 6000 up to 6000 Speed setting rotating pressing knob Speed display digital Timer 10 secon...

Reviews: