![IKA RW 47 D Operating Instructions Manual Download Page 14](http://html1.mh-extra.com/html/ika/rw-47-d/rw-47-d_operating-instructions-manual_624766014.webp)
The speed on the RW 47 D agitator is adjusted by means of
the rotary button (C).
You can read the set speed from the marking on the top of
the casing on the label attached to the motor support. (The
values are only valid for low agitator shaft lows).
You can select one of the three gear positions available (0,
I or II) using the toggle button (D) on the front of the
appliance. You may only switch to a different cog wheel
gear level when the appliance is switched off (motor
stopped). Otherwise this may cause damage to the gear
cogs. The arrow on the front plate indicates the direction of
the chuck.
Laboratory stirrer RW 47D must be secured to a stable stand
with four screws (e.g. R474 or R472 ) and wired according to
the wiring plan for proper use. The stirring bowl must always
be well secured for safety reasons. You must also ensure that
the mounting equipment (e.g. floor stand, telescope stand) is
set up in such a way that it is not liable to topple and does not
start to move during the stirring procedure. Accessories must
be assembled and wired according to the assembly
instructions.
If the stirring unit is operated in such a way that the appliance’s
EMERGENCY OFF switch ((B)-
fig. 2
) is not accessible, you
must add an another EMERGENCY OFF switch to the work
area so that it can be easily reached. Points of particular danger
(such as crushing, rotating parts, clamping screws for stand
legs and rollers etc.) are marked in
fig. 3
with an exclamation
mark. Please pay particular attention to these danger points
during operation!
14
RW47D102003
Speed display
Gear speed change
Commissioning
Fig. 3
RW 47 D_def 09.02.2004 14:54 Uhr Seite 14