![IKA RV 3 eco Instructions Manual Download Page 27](http://html1.mh-extra.com/html/ika/rv-3-eco/rv-3-eco_instructions-manual_624756027.webp)
27
Error sound | Causes | Effect | Solutions
Continuously beeping
Causes
› The speed has not been adjusted to 0 rpm when starting the device
› PCBA failure
Effect
› Motor can’t run
Solutions
› Turn the speed adjusting knob and set the speed to 0 rpm, after which you
may set the required speed value�
Alternatively beep 1 sec and stop 1 sec
Causes
› Overload
› Motor blocked
› Defective cable
› PCBA failure
Effect
› Motor stops running
Solutions
› Reduce the volume in the evaporator
› Contact the service department
Alternatively beep 0.5 sec and stop 1.5 sec
Causes
› Temperature inside the device is too high
› PCBA failure
Effect
› Motor stops running
Solutions
› Switch off the device
› Let the device cool down
› Switch on the device
› Contact the service department
Alternatively beep1.5 sec and stop 0.5 sec
Causes
› Overvoltage
› Undervoltage
› PCBA failure
› Desktop power supply failure
Effect
› Motor stops running
Solutions
› Use the original power supply to supply the device
› Contact the service department
Troubleshooting
�
Any malfunctions during operation will be identified by the error sound�
Proceed as follows in such cases:
› Turn off the device by using the main switch�
› Carry out corrective measures�
› Restart the device�
�
If the actions described fail to resolve the fault or the other kind of error occurred, then take one of the
following steps:
› contact the service department�
› send the device for repair, including a short description of the fault�