Connect the water hoses (
H
2
O
) to the condenser according to the counter-flow principle�
�
Install the vacuum connections to the condenser, condensate flask, vacuum controller with
valve and vacuum pump�
�
Always connect the vacuum hose (
vac
) to the condenser at the highest point to minimize
solvent losses during suctioning�
�
Use standard laboratory vacuum hoses with an inside diameter of 8 mm and a wall thickness
of 5 mm�
�
Refer to the chapter entitled “Operation“ in the operating instructions of heating bath�
/// Hose system
/// Heating bath
vac
H2O
H2O
pump