![IKA RCT basic safety control Operating Instructions Manual Download Page 27](http://html.mh-extra.com/html/ika/rct-basic-safety-control/rct-basic-safety-control_operating-instructions-manual_624732027.webp)
27
Warranty
In accordance with
IKA
warranty conditions, the warranty period
is 24 months. For claims under the warranty please contact your
local dealer. You may also send the machine direct to our factory,
enclosing the delivery invoice and giving reasons for the claim. You
will be liable for freight costs.
The warranty does not cover worn out parts, nor does it apply
to faults resulting from improper use, insufficient care or mainte-
nance not carried out in accordance with the instructions in this
operating manual.
Speed variation (no load, nominal voltage, at 1500 rpm, ambient temperature 25 °C)
%
± 2
Self-heating of the heating plate by max. stirring (RT:22 °C / duration:1 h)
K
+ 28
Stirred quantity max. (H
2
O)
ltr
20
Base plate
Dimensions
mm
Ø135
Material
Al-alloy
Heating
Heating power (-5%/+10%) at nominal voltage
W
600
Adjustment and display resolution
K
1
Surface temperature
°C
Ambient temperature ... 310
Heat control
rotating / pressing knob
Heating rate heating plate
K/min
6.5
Temperature sensor PT 1000 variation DIN EN 60751 Kl. A
K
≤ ± (0,15 +0,002xITI)
Temperature variation max. at 100 °C
K
± 1,5
Heating plate control hysteresis, no container, center of heating plate at 100 °C
K
± 5
Control hysteresis with temperature sensor PT 1000
Ü
with ETS-D5
Ü
with ETS-D6
Ü
K
°C
°C
± 1
± 0,5
± 0,2
Sensor in medium detection (Error 5)
yes
Adjustable safety circuit
Safety temperature limit (adjustable)
°C
(50 ... 360) ± 10
Ü
The control accuracy values given were determined using the following set-up: 500 ml water in 600 ml glass beaker, magnetic stirring bar 40, 600 rpm, 50 °C
Subject to technical changes!