3. Temperature sensor:
1) Calibration:
The internal and external temperature measurement can be calibrated and adjusted.
You can select 2-point calibration or 3-point calibration for internal and external measurement.
Calibration proceeding (example: 2-point calibration):
> Tempering
Temperature Sensor
Calibration
Reset Calibration
> Temperature Sensor
Calibration
Internal
External
> Calibration
Internal
2 Point Calibration
3 Point Calibration
> Calibration > Internal
2 Point Calibration
Temperature sensor calibration
at
Step: 1 / 2
5.00
oC
OK
> Calibration > Internal
2 Point Calibration
Temperature sensor calibration
at
Step: 1 / 2
30.00
oC
OK
> Calibration > Internal
2 Point Calibration
Temperature sensor calibration
at
Step: 1 / 2
30.00
oC
OK
> Calibration > Internal
2 Point Calibration
Target Temp:
30.00
oC
Actual Temp:
25.00
oC
Please wait...
Step: 1 / 2
> Calibration > Internal
2 Point Calibration
Target Temp:
30.00
oC
Actual Temp:
30.00
oC
Step: 1 / 2
OK
> Calibration > Internal
2 Point Calibration
Step: 1 / 2
Enter the temperature from the
measuring device
5.00
oC
OK
> Calibration > Internal
2 Point Calibration
Step: 1 / 2
Enter the temperature from the
measuring device
32.00
oC
OK
> Calibration > Internal
2 Point Calibration
Step: 1 / 2
Enter the temperature from the
measuring device
32.00
oC
OK
> Calibration > Internal
2 Point Calibration
Sensor calibrated successfully
OK
Fig. 27
2) Reset Calibration:
By “Reset Calibration”, the calibration value for the internal or the external temperature sensor will be deleted.
4. Notification:
1) ON:
This menu option allows you to activate/deactivate the “Notification“ function. A green check mark and a beep in-
forming when the target value (hysteresis) has been reached.
2) Hysteresis:
This menu option allows you to set the hysteresis from 0 to
±
2.0 K.
3) Sound:
Infinite: Beep until you press the “Back” button.
Beep:
Single beep.
PUMP
Limits:
In “Limits“ option, the user is allowed to set the maxi-
mum and minimum speed to the pump, and also the
maximum pressure.
TIMER
1. Set:
The user can set a target time (duration).
When device functions are started normally, this time is
displayed on the working screen. The device functions
stop automatically once this time has expired. The run-
ning time is then displayed again on the display.
Note:
To deactivate the target time specification, set
the target time to 00:00.
2. Display:
Activate the timer display on the main display (working
screen).