22
23
Reset to factory settings (r):
r 0:
Without resetting to factory settings�
r 1:
Resetting to factory settings�
Calibration of the other points is performed in the same way�
Transportation and storage
› Remove the power cord before transporting the device�
› The device must be protected against mechanical impact, vibrations, dust deposits and corrosive ambient
air during transportation and storage�
In addition, take care to ensure that the relative humidity does not exceed 80 %�
› The device must be completely emptied before transportation and storage�
1.
2.
1
2
›
Beware of the high dead weight of the device when transporting�
› Ensure that your fingers do not get crushed when setting down the device�
› The device can be transported:
- by hand / unaided (2 to 4 persons)�
- with a pallet truck�
Caution!
Beware of tipping over because of the high center of gravity when lift and move the device!
Confirm the value by pressing the “OK / Temp“ button� The first point calibration is finished now�
The screen displays the set value for the first point calibration and flashes�
Alarm and button tone (b):
b 0:
no alarm and button tone
b 1:
alarm and button tone is activated
Calibration and adjustment (C):
C 0:
Reset calibration
C 2:
2-point calibration
C 3:
3-point calibration�
Example: 2-point calibration:
Dip the temperature sensor of the reference measuring instrument into the bath fluid�
Select 2-point calibration with plus (+) button / minus (-) button and “OK / Temp“ button in
the menu�
Set the first point temperature (step 1, e�g� 10 °C) with plus (+) button / minus (-) button�
Confirm the setting by pressing “OK / Temp“ button� The unit now starts and controls to the
set value�
When the temperature has reached the set value and has become constant, the value in the
screen changes from the set temperature value to "0�0" and flashes�
Input the calibration value from the reference measuring instrument (e�g� 10�3 °C) with plus (+)
button / minus (-) button�
Note:
the difference between the input calibration value from reference measuring instrument
and set value is limited to less than +/-3 K�
Press the “OK / Temp“ button to start the 2-point calibration� The last calibration temperature
value flashes�