Reset to factory settings (r):

r 0: 

Without resetting to factory settings�

r 1: 

Resetting to factory settings�

Calibration of the other points is performed in the same way�

Transportation and storage

›  Remove the power cord before transporting the device� 
›  The device must be protected against mechanical impact, vibrations, dust deposits and corrosive ambient 

air during transportation and storage�

  In addition, take care to ensure that the relative humidity does not exceed 80 %�
›  The device must be completely emptied before transportation and storage�

1.

2.

1

2

› 

Beware of the high dead weight of the device when transporting�

›  Ensure that your fingers do not get crushed when setting down the device�
›  The device can be transported:

- by hand / unaided (2 to 4 persons)� 
- with a pallet truck� 

 Caution!

Beware of tipping over because of the high center of gravity when lift and move the device! 

Confirm the value by pressing the “OK / Temp“ button� The first point calibration is finished now�
The screen displays the set value for the first point calibration and flashes�

Alarm and button tone (b):

b 0: 

no alarm and button tone

b 1: 

alarm and button tone is activated

Calibration and adjustment (C):

C 0: 

Reset calibration

C 2: 

2-point calibration

C 3: 

3-point calibration�

Example: 2-point calibration:

Dip the temperature sensor of the reference measuring instrument into the bath fluid�
Select 2-point calibration with plus (+) button / minus (-) button and “OK / Temp“ button in 
the menu� 

Set the first point temperature (step 1, e�g� 10 °C) with plus (+) button / minus (-) button� 

Confirm the setting by pressing “OK / Temp“ button� The unit now starts and controls to the 
set value� 
When the temperature has reached the set value and has become constant, the value in the 
screen changes from the set temperature value to "0�0" and flashes�

Input the calibration value from the reference measuring instrument (e�g� 10�3 °C) with plus (+) 
button / minus (-) button� 

Note:

 the difference between the input calibration value from reference measuring instrument 

and set value is limited to less than +/-3 K�

Press the “OK / Temp“ button to start the 2-point calibration� The last calibration temperature 
value flashes�

Summary of Contents for RC 2 lite

Page 1: ...RC 2 lite...

Page 2: ...3 ENGLISH...

Page 3: ...11 Operation 16 Intended use 10 Technical Data 31 Maintenance and cleaning 28 Warranty 32 Pump performance curve 32 Error codes 29 Accessories 30 Operator panel and display 12 Fig 1 Interfaces and out...

Page 4: ...low the safety instructions Keep the operating instructions in a place where it can be accessed by everyone Ensure that only trained staff work with the device Follow the safety instructions guideline...

Page 5: ...1 g soda sodium carbonate Na2CO3 liter to reduce corrosive properties Don t use following fluids Untreated tap water Acids or bases Solutions with halides chlorides fluorides bromides iodides or sulf...

Page 6: ...er Do not damage the refrigerant circuit Insufficient ventilation may result in the formation of explosive mixtures Only use the device in well ventilated areas and keep all air vents open Flammable r...

Page 7: ...gs in the menu Change the menu value setting H OK Temp button Start Stop the tempering function Confirm the menu options B C D F G H A E Installation Place the unit on an even stable clean nonslip dry...

Page 8: ...14 15 Example Filling the device 1 Fluid level Low High 2 2x 3 Connecting to power 1 2 220 240 V 50 60 Hz 250 W IP 21 1...

Page 9: ...n Switching on Operation Notice Before commissioning make sure that the device has not been moved in one hour The socket used must be earthed fitted with earth contact Observe the ambient conditions t...

Page 10: ...Confirm the menu settings by pressing the OK Temp button H Menu navigation and structure The type of temperature PID control arithmetic A Menu Menu structure Default settings AUTO 1 Minimum temperatu...

Page 11: ...ility of the controller The differential time Td s is the derivative time and determines how strongly the rate of change of the control deviation affects the control variable Td compensates for rapid...

Page 12: ...d now The screen displays the set value for the first point calibration and flashes Alarm and button tone b b 0 no alarm and button tone b 1 alarm and button tone is activated Calibration and adjustme...

Page 13: ...2 Chemical composition Si Silicone oil EG Ethylene Glycol 3 Temperature range Minimum temperature Maximum temperature N Negative Temperature 4 Viscosiy Viscosiy at 25 C for Heating Fluid HF Viscosity...

Page 14: ...ldsoft provides a convenient tool for controlling the device and collecting data under MS Windows Command syntax and format The following applies to the command set Commands are generally sent from th...

Page 15: ...om the IKA website www ika com If you require servicing return the device in its original packaging Storage packaging is not sufficient Please also use suitable transport packaging Maintenance and cle...

Page 16: ...ensions W D H mm 225 x 430 x 385 Weight kg 24 5 Operation at a terrestrial altitude m max 2000 Note Refrigerant must be disposed of in accordance with local and national regulations Subject to technic...

Page 17: ...ll be liable for freight costs The warranty does not cover worn out parts nor does it apply to faults resulting from improper use insufficient care or maintenance not carried out in accordance with th...

Page 18: ...gland ika com MALAYSIA IKA Works Asia Sdn Bhd Phone 60 3 6099 5666 eMail sales lab ika my CHINA IKA Works Guangzhou Phone 86 20 8222 6771 eMail info ika cn POLAND IKA Poland Sp z o o Phone 48 22 201 9...

Reviews: