Interface and output

The device can be connected to a PC and operated with the 
laboratory software  labworld

soft

®  

through the RS 232 port 

(5), USB port (6) or USB port on the 

WiCo

Note: 

Please observe the system requirements together 

with the operating instructions and help section included 
with the software�

USB interface:

The Universal Serial Bus (USB) is a serial bus for connect-
ing the device to the PC� Equipped with USB devices can 
be connected to a PC during operation (hot plugging)� 
Connected devices and their properties are automatically 
recognized� Use the USB interface in conjunction with lab-
world

soft

®

 for operation in “Remote” mode and also to 

update the firmware�

USB device drivers:

First, download the latest driver for 

IKA

 devices with USB 

interface from: 

http://www.ika.com/ika/lws/download/usb-driver.zip.

 

Install the driver by running the setup file� Then connect 
the 

IKA

 device through the USB data cable to the PC� 

The data communication is via a virtual COM port� Con-
figuration, command syntax and commands of the virtual 
COM ports are as described in RS 232 port �

Device software update:

For device software update, visit 

IKA

 website 

www�ika�

com

 and enter the “Service” menu� Download and run 

the Firmware Update Tool�
Find and click the Firmware Update Tool in your PC after 
installation� Register your E-mail and password� 
Connect the device to your PC via the USB cable� Then, 
you can update the device software according to instruc-
tions of the Firmware Update Tool�

RS 232 interface:

Configuration
- The functions of the interface connections between the 

device and the automation system are chosen from the 
signals specified in EIA standard RS 232 in accordance 
with DIN 66 020 Part 1�

- For the electrical characteristics of the interface and the al-

location of signal status, standard RS 232 applies in accor-
dance with DIN 66 259 Part 1�

- Transmission procedure: asynchronous character trans-

mission in start-stop mode�

- Type of transmission: full duplex�
-  Character format: character representation in accordance 

with data format in DIN 66 022 for start-stop mode� 1 
start bit; 7 character bits; 1 parity bit (even); 1 stop bit�

- Transmission speed: 9600 bit/s�
- Data flow control: none
- Access procedure: data transfer from the device to the 

computer takes place only at the computer’s request�

Command syntax and format:

The following applies to the command set:
- Commands are generally sent from the computer (Mas-

ter) to the device (Slave)�

- The device sends only at the computer’s request� Even 

fault indications cannot be sent spontaneously from the 
device to the computer (automation system)�

-  Commands are transmitted in capital letters�
- Commands and parameters including successive parame-

ters are separated by at least one space (Code: hex 0x20)�

- Each individual command (incl� parameters and data) 

and each response are terminated with Blank CR LF 
(Code: hex 0x0d hex 0x0A) and have a maximum length 
of 80 characters�

- The decimal separator in a number is a dot (Code: hex 

0x2E)�

The above details correspond as far as possible to the recommendations of the NAMUR working party (NAMUR recom-
mendations for the design of electrical plug connections for analogue and digital signal transmission on individual items 
of laboratory control equipment, rev� 1�1)�
The NAMUR commands and the additional specific

 IKA

 co mmands serve only as low level commands for communica-

tion between the device and the PC� With a suitable terminal or communications programme these commands can be 
transmitted directly to the circulator equipment� The 

IKA

 software package, Labworld

soft

®

, provides a convenient tool 

for controlling circulating equipment and collecting data under MS Windows, and includes graphical entry features, for 
pump motor speed ramps for example�

Commands:
Commands

Function

IN_PV_1

Read the external actual temperature

IN_PV_2

Read the internal actual temperature

IN_PV_4

Read the pump actual speed

IN_SP_1

Read the internal setting temperature (if 0: internal control)
Read the external setting temperature (if 1: external control)

IN_SP_4

Read the pump setting speed

Summary of Contents for RC 2 GREEN control

Page 1: ...Operating instructions EN IKA RC 2 GREEN control...

Page 2: ...ig 1 Item Designation 1 Filling opening lid 2 Wireless Controller WiCo 3 Handle 4 Venting grid 5 RS 232 port 6 USB port 7 External temperature sensor socket 8 Pump connector IN 9 Pump connector OUT 10...

Page 3: ...g battery into the WiCo 9 Mounting the WiCo to the station 9 Filling and draining 9 Fluids Standard information for IKA fluid 11 Moving the device 12 Charging the Battery Pack RB 1 rechargeable batter...

Page 4: ...nse exempt RSS standard s Operation is subject to the following two conditions 1 This device may not cause interference 2 This device must accept any interference including interference that may cause...

Page 5: ...on the device again to ensure safe operation for example to keep foreign ob jects and liquids etc from getting into the device The device may only be used as prescribed and as de scribed in these ope...

Page 6: ...OFF key and the Power switch Please note the following safety instructions for the battery pack RB 1 recharge able battery Keep the battery pack out of reach of children at all times Store the batter...

Page 7: ...k between the WiCo and the laboratory device first check whether your region is in cluded in the radio communication approval for the de vice If it is not remote control can also be performed using a...

Page 8: ...he cooling system Connecting the tubings Unscrew the union nuts and stoppers using a wrench AF 19 from the pump connector IN 8 and OUT 9 Connect the hoses for circulating the external system to the pu...

Page 9: ...w cap Inserting battery into the WiCo Model WiCo CBC 5 LPS IP 40 Model WiCo CBC 5 LPS IP 40 Model WiCo CBC 5 LPS IP 40 Note notch upward Model WiCo CBC 5 LPS IP 40 Model WiCo CBC 5 LPS IP 40 Fig 7 Scr...

Page 10: ...ckwise to the stop position see Fig 10 Note Please note information in section Commissioning Connect the power plug and turn on the device with power switch 10 The low level icon appears on the displa...

Page 11: ...perature Maximum temperature N Negative Temperature 4 Viscosiy Viscosiy at 25 C for Heating Fluid HF Viscosity at 20 C for Cooling Fluid CF Viscosity at 25 C for Universal Fluid UF Dynamic viscosity m...

Page 12: ...ss head clamp H 38 Holding rod IKA RC 2 GREENcontrol IKA RC 2 GREEN control IKA RC 2 GREENcontrol Moving the device Empty all fluid in the bath before move device from one place to other place The dev...

Page 13: ...gh Low Fluid level 2 GREEN Fig 16 A F D E B C Item Designation Function L On Off button Switch the WiCo on and off M int ext button Switch between the internal and external temperature display and con...

Page 14: ...nce The refrigerating function may delay to start at maximum 2 minutes In working status press the right knob R to stop opera tion of the pump The refrigerating function and the pump stops Note In sta...

Page 15: ...and the station The screen below shows the most significant symbols on the working screen Bluetooth This symbol means the station and the WiCo are com municating via Bluetooth The symbol no longer app...

Page 16: ...he menu by turning the right rotating pressing knob R to the right or left Open the menu item by pressing the right rotating pressing knob R Turn the rotating pressing knob R to select the desired men...

Page 17: ...omized 1 Customized 2 PUMP Limits Speed Pressure Maximum Maximum Minimum 4000 rpm 2000 rpm 0 5 bar TIMER Set Display 00 00 hh mm MODE A C B D activated GRAPH Automatic Manual Axis Assignment Set Tempe...

Page 18: ...e difference between the external system and internal bath during the heating up and cooling down process This can protects sensitive device like the glass reactors from thermal shock e g glass break...

Page 19: ...Step 1 2 Enter the temperature from the measuring device 32 00 OC OK Calibration Internal 2 Point Calibration Step 1 2 Enter the temperature from the measuring device 32 00 OC OK Calibration Internal...

Page 20: ...d or set PROGRAMS Under programs 10 user defined temperature time pro files can be created A program can consist of up to 10 segments Once a program has been selected the following options are availab...

Page 21: ...button re sets all the system settings to the original standard values set at dispatch from the factory see Menu structure 5 Communication The Device name menu option allows you to edit the device na...

Page 22: ...top mode Type of transmission full duplex Character format character representation in accordance with data format in DIN 66 022 for start stop mode 1 start bit 7 character bits 1 parity bit even 1 st...

Page 23: ...1 and set the time for the watchdog to n 20 1500 seconds Echo of the Watchdog time During a WD1 event the cooling and pump functions are switched off and message PC 1 is display This command needs to...

Page 24: ...o A USB micro B USB micro A Connection the device to PC USB micro B USB A USB micro B USB A RS 232 USB or USB micro B USB A Fig 32 USB cable RS 232 USB or USB micro B USB A 9 pin RS 232 9 pin RS 232 A...

Page 25: ...alcohol Wear protective gloves while cleaning the instruments Electrical instruments may not be placed in the cleansing agent for the purpose of cleaning Do not allow moisture to get into the instrume...

Page 26: ...impeller is blocked Error 06 Pump off Refrigerating off Too low liquid level Check the liquid level and buoyage Error 10 Pump off Refrigerating off PC communication failure Check communication cable E...

Page 27: ...h 12 H SI 8 Silicone hose nominal width 8 H SI 12 Silicone hose nominal width 12 H PUR 8 PUR tube nominal width 8 mm H PUR 12 PUR tube nominal width 12 mm H FKM 8 FKM tube nominal width 8 mm H FKM 12...

Page 28: ...Knob on WiCo Temperature setting resolution K 0 1 Temperature display TFT LCD on WiCo Temperature display resolution K 0 01 Cooling capacity according to DIN 12876 at 4000 rpm 20 C 10 C 0 C 10 C 20 C...

Page 29: ...ranty conditions the warranty period is 24 months For claims under the warranty please contact your local dealer You may also send the machine direct to our factory enclosing the delivery invoice and...

Page 30: ...es england ika com MALAYSIA IKA Works Asia Sdn Bhd Phone 60 3 6099 5666 eMail sales lab ika my CHINA IKA Works Guangzhou Phone 86 20 8222 6771 eMail info ika cn POLAND IKA Poland Sp z o o Phone 48 22...

Reviews: