Commissioning and operating

Note: 

Before commissioning, make sure that the device has 

not been moved for one hour!

Check whether the voltage given on the type plate cor-
responds to the available power voltage� 

The socket used must be earthed 
(fitted with earth contact)� 

If these conditions have been met, the machine is ready 
for operation when the power plug is plugged in� 
If these conditions are not met, safe operation is not guar-
anteed and the machine could be damaged�

Observe  the ambient conditions (temperature, humidity, 
etc�) listed under “Technical Data”�

After switching on the power switch (10) of the 

station

the screen of the 

WiCo

 displays the device designation 

and the software version after a beep� 

After several seconds, screen display the information of

 

WiCo

� 

IKA

®

Circulator 

RC Control

Version 1.2.008

Then the working screen appears and the device is ready 
for operation�

Information

Internal Max: 

40 °C

Internal Min: 

5 °C

External Max: 

40 °C

External Min: 

5 °C

Fluid: H

2

O (Water)

Change the temperature setting with left knob (P)�
Change the pump speed setting with the right knob (R)�

In standby status, activate the refrigerating function by 
pressing the left knob (P)� The pump and fan starts to run 
at once� The refrigerating function may delay to start at 
maximum 2 minutes�

In working status, press the right knob (R) to stop opera-
tion of the pump� The refrigerating function and the pump 
stops�

Note:

 In standby status,

 

press the right knob (R) to start 

the pump function, the refrigerating function will not be 
activated�
In working status, press the left knob (P) to stop the refrig-
erating function, the pump keep running�

When the 

WiCo

 is switched on without being connected 

to the 

station

, the green LED bar (B) and the green Blue-

tooth

®

 LED (F) on 

station 

lights up� 

The control elements of the 

WiCo

 allow locked by press-

ing the lock button (V) about 2 seconds, so no accidental 
changes during operation are possible (lock symbol 

 

appears in the display)�
By pressing lock button (V) about 2 seconds again, the 
control elements are released (lock symbol 

 disappears 

from the display)�

Note:

 In an emergency, the device function can be turned 

off by pressing the “

safe STOP

” button (D) at the front of 

station

� The LED bar (B) changed into red color�

A message appears in the display indicates that the

 station

 

was forced off� To restart, press the power switch (10) off 
and on again�
If the Bluetooth

®

 function of the 

WiCo

 is activated, the 

Bluetooth

®

 icon   appears on the screen and the user 

can search the 

WiCo

 by pressing the searching button (E)� 

Then, a beep is heard�

Fig. 18

Fig. 19

Fig. 20

A

0�00

int °C

ext °C

int

5�00

2000

set °C

Pump rpm

Summary of Contents for RC 2 GREEN control

Page 1: ...Operating instructions EN IKA RC 2 GREEN control...

Page 2: ...ig 1 Item Designation 1 Filling opening lid 2 Wireless Controller WiCo 3 Handle 4 Venting grid 5 RS 232 port 6 USB port 7 External temperature sensor socket 8 Pump connector IN 9 Pump connector OUT 10...

Page 3: ...g battery into the WiCo 9 Mounting the WiCo to the station 9 Filling and draining 9 Fluids Standard information for IKA fluid 11 Moving the device 12 Charging the Battery Pack RB 1 rechargeable batter...

Page 4: ...nse exempt RSS standard s Operation is subject to the following two conditions 1 This device may not cause interference 2 This device must accept any interference including interference that may cause...

Page 5: ...on the device again to ensure safe operation for example to keep foreign ob jects and liquids etc from getting into the device The device may only be used as prescribed and as de scribed in these ope...

Page 6: ...OFF key and the Power switch Please note the following safety instructions for the battery pack RB 1 recharge able battery Keep the battery pack out of reach of children at all times Store the batter...

Page 7: ...k between the WiCo and the laboratory device first check whether your region is in cluded in the radio communication approval for the de vice If it is not remote control can also be performed using a...

Page 8: ...he cooling system Connecting the tubings Unscrew the union nuts and stoppers using a wrench AF 19 from the pump connector IN 8 and OUT 9 Connect the hoses for circulating the external system to the pu...

Page 9: ...w cap Inserting battery into the WiCo Model WiCo CBC 5 LPS IP 40 Model WiCo CBC 5 LPS IP 40 Model WiCo CBC 5 LPS IP 40 Note notch upward Model WiCo CBC 5 LPS IP 40 Model WiCo CBC 5 LPS IP 40 Fig 7 Scr...

Page 10: ...ckwise to the stop position see Fig 10 Note Please note information in section Commissioning Connect the power plug and turn on the device with power switch 10 The low level icon appears on the displa...

Page 11: ...perature Maximum temperature N Negative Temperature 4 Viscosiy Viscosiy at 25 C for Heating Fluid HF Viscosity at 20 C for Cooling Fluid CF Viscosity at 25 C for Universal Fluid UF Dynamic viscosity m...

Page 12: ...ss head clamp H 38 Holding rod IKA RC 2 GREENcontrol IKA RC 2 GREEN control IKA RC 2 GREENcontrol Moving the device Empty all fluid in the bath before move device from one place to other place The dev...

Page 13: ...gh Low Fluid level 2 GREEN Fig 16 A F D E B C Item Designation Function L On Off button Switch the WiCo on and off M int ext button Switch between the internal and external temperature display and con...

Page 14: ...nce The refrigerating function may delay to start at maximum 2 minutes In working status press the right knob R to stop opera tion of the pump The refrigerating function and the pump stops Note In sta...

Page 15: ...and the station The screen below shows the most significant symbols on the working screen Bluetooth This symbol means the station and the WiCo are com municating via Bluetooth The symbol no longer app...

Page 16: ...he menu by turning the right rotating pressing knob R to the right or left Open the menu item by pressing the right rotating pressing knob R Turn the rotating pressing knob R to select the desired men...

Page 17: ...omized 1 Customized 2 PUMP Limits Speed Pressure Maximum Maximum Minimum 4000 rpm 2000 rpm 0 5 bar TIMER Set Display 00 00 hh mm MODE A C B D activated GRAPH Automatic Manual Axis Assignment Set Tempe...

Page 18: ...e difference between the external system and internal bath during the heating up and cooling down process This can protects sensitive device like the glass reactors from thermal shock e g glass break...

Page 19: ...Step 1 2 Enter the temperature from the measuring device 32 00 OC OK Calibration Internal 2 Point Calibration Step 1 2 Enter the temperature from the measuring device 32 00 OC OK Calibration Internal...

Page 20: ...d or set PROGRAMS Under programs 10 user defined temperature time pro files can be created A program can consist of up to 10 segments Once a program has been selected the following options are availab...

Page 21: ...button re sets all the system settings to the original standard values set at dispatch from the factory see Menu structure 5 Communication The Device name menu option allows you to edit the device na...

Page 22: ...top mode Type of transmission full duplex Character format character representation in accordance with data format in DIN 66 022 for start stop mode 1 start bit 7 character bits 1 parity bit even 1 st...

Page 23: ...1 and set the time for the watchdog to n 20 1500 seconds Echo of the Watchdog time During a WD1 event the cooling and pump functions are switched off and message PC 1 is display This command needs to...

Page 24: ...o A USB micro B USB micro A Connection the device to PC USB micro B USB A USB micro B USB A RS 232 USB or USB micro B USB A Fig 32 USB cable RS 232 USB or USB micro B USB A 9 pin RS 232 9 pin RS 232 A...

Page 25: ...alcohol Wear protective gloves while cleaning the instruments Electrical instruments may not be placed in the cleansing agent for the purpose of cleaning Do not allow moisture to get into the instrume...

Page 26: ...impeller is blocked Error 06 Pump off Refrigerating off Too low liquid level Check the liquid level and buoyage Error 10 Pump off Refrigerating off PC communication failure Check communication cable E...

Page 27: ...h 12 H SI 8 Silicone hose nominal width 8 H SI 12 Silicone hose nominal width 12 H PUR 8 PUR tube nominal width 8 mm H PUR 12 PUR tube nominal width 12 mm H FKM 8 FKM tube nominal width 8 mm H FKM 12...

Page 28: ...Knob on WiCo Temperature setting resolution K 0 1 Temperature display TFT LCD on WiCo Temperature display resolution K 0 01 Cooling capacity according to DIN 12876 at 4000 rpm 20 C 10 C 0 C 10 C 20 C...

Page 29: ...ranty conditions the warranty period is 24 months For claims under the warranty please contact your local dealer You may also send the machine direct to our factory enclosing the delivery invoice and...

Page 30: ...es england ika com MALAYSIA IKA Works Asia Sdn Bhd Phone 60 3 6099 5666 eMail sales lab ika my CHINA IKA Works Guangzhou Phone 86 20 8222 6771 eMail info ika cn POLAND IKA Poland Sp z o o Phone 48 22...

Reviews: