23
Any malfunctions during operation will be identified by an error message on the display (H), see fig. 6.
Proceed as follows in such cases:
☞
Switch off device using the main switch (B)
☞
Carry out corrective measures
☞
Restart device
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.
Error message HB 10 digital/control
Error message
Cause
Effect
Correction
E 2
- In remote operation (PC), no communication
Heating off
- Switch off the device
between RV 10 and HB 10
- Clean the IR interface or
- IR connection to RV 10 disconnected
remove objects from around the IR interface
- Switch on the device
E 3
- Temperature inside the device is too high
Heating off
- Switch off the device
- Let the device cool down
- Switch on the device
E 9
- Fault in storing the safety circuit set temperatures-
Heating off
- Switch off the device
- Memory chip (EPROM) defective
- Let the device cool down
- Switch on the device
E 21
- Safety relay does not open
Heating off
- Switch off the device and switch on again
E 23
- Adjustable safety circuit defective
Heating off
- Switch off the device and switch on again
- Allow the device to cool down
- Switch on the device
E 24
- Set safety temperature exceeded
Heating off
- Switch off the device and switch on again
- Allow the device to cool down
- Switch on the device again
- Check the safety temperature setting
E 25
- The heater control circuit switch (TRIAC) is faulty
Heating off
- Switch off the device and switch on again
The heater or the supply line is disconnected
E 26
- Dry running
Heating off
- Let the device cool down
- Fill up the tempering medium
- Switch off the device and switch on again
E 27
- Error in calibration
Heating off
- Switch off the device and switch on again
E 28
- Breakage of the controller sensor
Heating off
- Switch off the device and switch on again
E 29
- Short-circuit in the safety sensor
Heating off
- Switch off the device and switch on again
E 30
- Short-circuit in the controller sensor
Heating off
- Switch off the device and switch on again
3946600_HB10bdc_032009:3345300_VG3_042005_kvmb_1 28.04.2009 9:23 Uhr Seite 23
Summary of Contents for HB 10
Page 37: ...37 7 3946600_HB10bdc_032009 3345300_VG3_042005_kvmb_1 28 04 2009 9 23 Uhr Seite 57...
Page 39: ...39 HB 10 2 HB 10 1 HB 10 1 HB 10 HB 10 HB 10 RV 10 IKA 1 2 5 HB 10 G D C E F Fig 12 Fig 12...
Page 44: ...44 HB 10 Error Error Error...
Page 45: ...45 HB 10 6 H B E 2 RV 10 E 3 E 9 EPROM E 21 E 23 E 24 E 25 E 26 E 27 E 28 E 29 E 30...
Page 46: ...46 IKA www ika cn HB 10 1 HB 10 2 HB 10 1 IKA...