22
Any malfunctions during operation will be identified by lighting up the red signal lamp “Error”.
Proceed as follows in such cases:
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the device for repair, including a short description of the fault.
Error message HB 10 basic
Error message
Cause
Effect
Correction
”Error” signal lamp
- Safety circuit temperature set too low
Device off
- Let device cool down
lights up
- Dry run
- Reset
- Controller set higher than safety circuit
- Check settings and correct if necessary
- Top up tempering medium
”Error” signal lamp
- Faulty controller
Device defect - Contact the service department.
lights up again after
reset, also in case of
max. adjusted
safety temperature
limit
3946600_HB10bdc_032009:3345300_VG3_042005_kvmb_1 28.04.2009 9:23 Uhr Seite 22